CareFusion AVEA

Exhalation Valve Failure

On this page

Asset Type

Ventilator

Manufacturer

CareFusion

Model

AVEA

What This Guide Helps With

This guide assists in troubleshooting exhalation valve failures on a CareFusion AVEA ventilator. Exhalation valve issues can prevent proper expiration of patient breaths, trigger alarms, and compromise ventilation.

Step-by-Step Troubleshooting

Ensure Patient Safety

Inspect the Exhalation Valve Assembly

Check Valve Alignment and Seating

Verify Circuit Connections

Inspect for Condensation or Moisture

Check Valve Sensor Functionality (If Equipped)

Power Cycle the Ventilator

Replace Disposable Components

If the Problem Persists

External causes have likely been ruled out.

Label the ventilator Out of Service and remove it from clinical use.

Send the device for bench evaluation or repair by qualified service personnel.

Knowing when to stop prevents patient risk and further equipment damage.

Clinical Use Tip

Never troubleshoot the exhalation valve while a patient is connected.

Always secure an alternative ventilation method before manipulating the valve or circuit.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
Ventilator alarmed with exhalation valve failure during routine use.

Cause

What was observed during troubleshooting.

Example:
Valve found obstructed with condensation and partially misaligned in expiratory limb.

Resolution

What action was taken.

Example:
Cleared obstruction, reseated valve, replaced circuit, and verified proper function; ventilator returned to service.

Helpful Details to Include (If Known)

Final Thought

Exhalation valve troubleshooting relies on careful observation, logical stepwise checks, and patient safety. Always rule out external causes before assuming internal failure, and document all findings clearly. Escalation to repair is the correct step when the valve issue cannot be resolved externally.

That is successful troubleshooting.