Carestream DRX-Evolution

Image Acquisition Failure or Exam Not Completing

On this page

Asset Type

Radiographic System

Manufacturer

Carestream

Model

DRX-Evolution

What This Guide Helps With

Image acquisition failure, incomplete exams, detector communication issues, workstation problems, or workflow interruptions on a Carestream DRX-Evolution system.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot the DRX-Evolution while the patient is positioned for an active exam unless the issue is minor and does not delay care.

Action:

Expected outcome: Patient care continues without relying on a system that may not complete the exam.

Why it matters: Image acquisition failures can delay diagnosis, increase repeat positioning, or lead to unnecessary repeat exposure attempts.

2. Verify the Reported Failure

Confirm exactly where the exam is failing.

Action:

Expected outcome: The failure point is identified before replacing accessories or escalating.

If the issue is resolved after correcting workflow or restarting the exam, stop.

3. Check for Active Error Messages or System Prompts

Look for visible messages on the acquisition workstation, generator console, detector status area, or room control interface.

Action:

Expected outcome: Any displayed fault is documented and used to guide the next step.

Why it matters: Many image acquisition failures are caused by incomplete workflow steps rather than hardware failure.

4. Confirm the Correct Patient and Exam Are Selected

Verify that the correct patient, accession number, body part, projection, and detector are selected before exposure.

Action:

Expected outcome: The system is ready to receive the exposure under the correct exam.

If the technologist had the wrong exam, projection, or detector selected and correcting it allows acquisition, stop.

5. Check Detector Status and Connection

Verify that the DRX detector is available, powered, connected, and assigned to the room.

Action:

Expected outcome: Detector status changes to ready, connected, or available.

If the detector reconnects and the exam completes normally, stop.

6. Inspect Detector Battery and Charging Contacts

Check for a low battery, loose battery, damaged battery latch, or dirty/damaged charging contacts.

Action:

Expected outcome: Detector power is stable and the detector remains connected during acquisition.

If a battery swap resolves the issue, document the failed battery and remove it from use.

7. Check Room Network and Workstation Connectivity

Confirm the acquisition workstation has network communication and is not disconnected from required hospital or imaging systems.

Action:

Expected outcome: The workstation has stable network communication.

Why it matters: Some exams may acquire locally but fail to complete, close, or transfer if network communication is interrupted.

8. Restart the Acquisition Application

If the workstation appears frozen, delayed, or stuck in the exam workflow, restart the acquisition software if safe and allowed by site procedure.

Action:

Expected outcome: The software responds normally and allows the exam workflow to continue.

If the application restart resolves the issue, document the software lockup and final status.

9. Power Cycle the Workstation and Room System

If the system remains locked, not ready, or unable to complete acquisition, perform a controlled restart.

Action:

Expected outcome: The workstation, detector, generator, and room system return to normal ready status.

If the system completes a test acquisition after restart, stop.

10. Verify Generator and Exposure Readiness

Check whether the acquisition failure is actually caused by the generator not completing exposure.

Action:

Expected outcome: The system reaches exposure-ready status before image acquisition is attempted.

Why it matters: If the generator never completes exposure, the workstation may appear to have an image acquisition problem even though the exposure was inhibited.

11. Check Bucky, Table, Wall Stand, and Detector Selection

Confirm the selected receptor matches the physical imaging setup.

Action:

Expected outcome: The system is not waiting for a different image receptor than the one being used.

If correcting receptor selection resolves the failed acquisition, stop.

12. Perform a Non-Patient Test Image if Permitted

After patient safety is addressed, perform a controlled test exposure using site-approved phantom or QA process.

Action:

Expected outcome: The acquisition chain is verified from exposure through image display and completion.

Do not perform unnecessary radiation exposures outside approved testing procedures.

13. Check for Repeat Failure Pattern

Determine whether the issue is isolated or repeatable.

Action:

Expected outcome: The failure pattern helps determine whether the issue is detector-related, workstation-related, generator-related, network-related, or workflow-related.

14. Remove Failed Accessories from Service

If one detector, battery, cable, charger, or network cable is clearly associated with the failure, remove that item from use.

Action:

Expected outcome: The room is returned to service only if the remaining configuration passes functional testing.

If the issue follows the accessory, send the accessory for repair or replacement.

If the Problem Persists

If the Carestream DRX-Evolution still cannot acquire images or complete exams after checking patient safety, exam workflow, detector status, battery condition, receptor selection, workstation behavior, network connection, and generator readiness, common external causes have been ruled out.

The system should be removed from service, labeled Out of Service, and sent for repair, vendor support, or bench/system evaluation.

Do not continue repeated exposure attempts or workflow resets if the failure is repeatable. Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Do not troubleshoot image acquisition failures while a patient is waiting in position unless the correction is immediate and safe. Move the patient to another room or imaging device first if care could be delayed. Maintain therapy and imaging continuity by using a backup device or approved alternate workflow before continuing troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Technologist reported the Carestream DRX-Evolution would not complete the exam after exposure, and the acquired image did not display on the workstation."

Cause

What was observed during troubleshooting.

Example:
"Found the wireless DRX detector intermittently disconnecting due to a depleted detector battery."

Resolution

What action was taken.

Example:
"Replaced the detector battery with a charged known-good battery, confirmed detector connection, completed a test acquisition, and returned the room to service."

Helpful Details to Include (If Known)

Final Thought

Image acquisition failures should be handled logically from the outside in. Protect the patient first, verify workflow and detector status, check power and communication, then confirm whether the generator actually completed exposure. If the problem remains repeatable after basic checks, remove the DRX-Evolution from service and escalate with clear CCR documentation.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide