Carestream DRX-Evolution

PACS Send Failure or Worklist Connectivity Issue

On this page

Asset Type

Radiographic System

Manufacturer

Carestream

Model

DRX-Evolution

What This Guide Helps With

Troubleshooting DRX-Evolution PACS send failures, missing worklist updates, network issues, destination errors, or failed image transfer.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Do not troubleshoot network transfer issues during active imaging if it delays patient care.

Action:

Expected outcome: Patient care continues safely while image transfer or worklist troubleshooting is handled separately.

If this resolves the workflow concern, stop.

Verify the Reported Problem

Confirm whether the issue is with PACS sending, worklist retrieval, or both.

Action:

Expected outcome: The failure type is clearly identified before troubleshooting continues.

Check for Facility-Wide PACS or Network Problems

A room issue should not be assumed until broader system problems are ruled out.

Action:

Expected outcome: You determine whether the DRX-Evolution room is isolated or part of a larger outage.

If multiple systems are affected, escalate to IT/imaging informatics and stop local equipment troubleshooting.

Confirm the System Is on the Correct Network

PACS and worklist communication depend on the acquisition workstation being connected to the correct hospital network.

Action:

Expected outcome: The workstation has a valid physical or wireless network connection.

If reconnecting the network restores worklist or PACS transfer, document the connection issue and stop.

Restart the Acquisition Workstation Only If Safe

A software or communication service may recover after a controlled restart.

Action:

Expected outcome: Communication services reload and normal DICOM transfer or worklist query returns.

If the restart resolves the issue, stop.

Check the Worklist Search Criteria

Missing worklist entries can be caused by search filters rather than system failure.

Action:

Expected outcome: The correct patient order appears if the issue was caused by filtering or incomplete order entry.

If the worklist appears after correcting filters, stop.

Verify DICOM Destination Selection

Images may fail to send if the wrong PACS destination or archive route is selected.

Action:

Expected outcome: Images are routed to the correct configured PACS destination.

If sending succeeds after selecting the correct destination, stop.

Review the Send Queue or Failed Jobs

The queue can show whether images are waiting, failed, rejected, or actively transferring.

Action:

Expected outcome: You identify whether the issue is a single failed study or a broader transfer failure.

Confirm Patient and Exam Data Are Valid

PACS may reject studies with missing or mismatched patient, accession, or exam information.

Action:

Expected outcome: The study contains valid demographic and order information for PACS acceptance.

If correcting demographics or order details allows the image to send, stop.

Check for Local Storage or Queue Backup

A full local drive or overloaded queue can prevent normal image transfer.

Action:

Expected outcome: Storage or queue limitations are identified without risking image loss.

Test Basic Network Function If Available

Clinical Engineering should verify simple network indicators before assuming equipment failure.

Action:

Expected outcome: Basic network communication is either confirmed or escalated to IT.

Escalate Configuration or Server Issues

DICOM AE titles, IP addresses, ports, and server-side routing are usually controlled by imaging informatics, IT, or vendor service.

Action:

Expected outcome: The correct support team can troubleshoot DICOM routing, PACS, RIS, firewall, or workstation configuration.

If the Problem Persists

If PACS send failure or worklist connectivity problems continue after checking patient safety, network connection, worklist filters, destination selection, send queue status, patient demographics, and basic connectivity, the likely cause is deeper software, DICOM configuration, network routing, or workstation failure.

The DRX-Evolution room should be:

Knowing when to stop is proper troubleshooting. Do not perform unauthorized DICOM configuration changes or risk image loss by deleting studies without confirmation.

Clinical Use Tip

Do not troubleshoot PACS transfer problems on an active patient if it delays care. Finish the exam safely, preserve acquired images, and move the patient to another room if imaging must continue.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Radiology reported that the Carestream DRX-Evolution could not send completed studies to PACS and the modality worklist was not updating."

Cause

What was observed during troubleshooting.

Example:
"Found network cable loose at the acquisition workstation, causing intermittent PACS and worklist communication failure."

Resolution

What action was taken.

Example:
"Reseated network cable, verified link lights, restarted the workstation, refreshed worklist, and confirmed test study successfully transferred to PACS."

Helpful Details to Include (If Known)

Final Thought

PACS and worklist issues should be approached logically before assuming a system failure. Start with patient safety, confirm whether the issue is local or facility-wide, check physical network connections, verify filters and destinations, and protect patient images from accidental loss. Good documentation helps Clinical Engineering, IT, imaging informatics, and vendor service quickly identify whether the issue is workflow, network, configuration, or equipment related.

That is successful troubleshooting.