Carestream DRX-Revolution

Image not sending to PACS or worklist not updating

On this page

Asset Type

Mobile X-Ray System

Manufacturer

Carestream

Model

DRX-Revolution

What This Guide Helps With

Troubleshooting PACS send failures, missing worklist updates, network issues, wireless connection problems, and image transfer delays on a DRX-Revolution.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot image transfer or worklist issues during an active patient exam unless the clinical workflow is already complete.

Action:

Expected outcome: Patient care continues safely, and completed images are protected from accidental loss.

Why it matters: PACS and worklist issues are usually workflow or connectivity problems, but mishandling stored images can cause repeat imaging or documentation problems.

2. Verify the Reported Problem

Confirm whether the issue is with image sending, worklist updates, or both.

Action:

Ask the technologist what specifically failed:

Check the DRX-Revolution display for network, DICOM, PACS, worklist, or send queue messages.

Expected outcome: The failure type is clearly identified before troubleshooting.

If the issue is only one patient or accession, focus on order/worklist data first. If all exams fail, focus on network or DICOM connectivity.

3. Confirm the Exam Was Completed Correctly

Make sure the image is available and ready to send.

Action:

Expected outcome: The image should be present locally and available for DICOM transfer.

If the image was never completed or is still held locally, complete the exam workflow and attempt to resend. If successful, stop.

4. Check the Send Queue

Look for images stuck in the transmission queue.

Action:

Open the local send queue, network queue, or image transmission status area.

Look for:

Attempt a manual resend if the image is safely stored and the system allows it.

Expected outcome: Images should either send successfully or produce a clear error that helps identify the cause.

If manual resend works, document the temporary transfer delay and stop.

5. Verify Wireless Network Connection

Mobile X-ray systems commonly fail to send images when wireless coverage is weak or disconnected.

Action:

Expected outcome: The unit should show a stable network connection.

If the unit reconnects and images send successfully, the likely cause was wireless coverage or roaming. Stop and document the location.

6. Check Whether the Issue Is Location-Specific

Determine if the problem follows the unit or the wireless area.

Action:

Expected outcome: The issue should be narrowed to the unit, network area, or hospital wireless infrastructure.

If multiple units fail in the same location, escalate to IT/network support rather than replacing imaging hardware.

7. Confirm Network Login or Authentication Status

Some systems may appear powered on but not fully authenticated to the hospital network.

Action:

Expected outcome: The unit should be fully logged into the system and connected to hospital network resources.

If login or authentication restores PACS/worklist function, stop and document the access issue.

8. Verify Worklist Search Criteria

A worklist may fail to update because the search filters do not match the expected order.

Action:

Check the worklist query fields, including:

Clear unnecessary filters and refresh the worklist.

Search by accession number if available.

Expected outcome: The correct order should appear if the worklist server is reachable and the query matches the order.

If the order appears after changing filters, the device is likely functioning normally. Stop and document the filter/search issue.

9. Confirm the Order Exists in RIS/EMR

Do not assume the DRX-Revolution is failing if the order was not placed or released correctly.

Action:

Expected outcome: A valid order should exist before the worklist can retrieve it.

If the order is missing or incorrect in the RIS/EMR, the issue is not a device failure. Stop and document that the order/worklist source requires correction.

10. Check PACS and Worklist Destination Availability

The device may be working, but the receiving system may be down or unavailable.

Action:

Expected outcome: Hospital imaging systems should confirm whether this is isolated to the DRX-Revolution or part of a larger outage.

If there is a known PACS/worklist outage, stop device troubleshooting and document the external system issue.

11. Verify Correct PACS Destination Selection

Some systems may have multiple send destinations.

Action:

Expected outcome: Images should be sent to the intended production PACS destination.

If the wrong destination was selected, resend to the correct destination and stop if successful.

12. Confirm Date and Time Accuracy

Incorrect system date or time can affect worklist queries, image timestamps, and DICOM routing.

Action:

Expected outcome: The unit should match hospital time closely.

If the date/time is incorrect and correcting it restores worklist or PACS behavior, stop and document the finding.

13. Reboot the System Only After Image Status Is Known

A reboot may restore network services, but it should not be the first action if unsent images are present.

Action:

Expected outcome: Temporary software or network service issues may clear after reboot.

If reboot restores PACS or worklist function, document the fault as intermittent communication/software state issue and stop.

14. Check for Physical Network Options if Available

If the unit supports wired network connection or docking, use it to separate wireless issues from DICOM configuration problems.

Action:

Expected outcome: If wired network works but wireless does not, the issue is likely wireless connectivity, configuration, antenna, or network coverage related.

If wired transfer works, document the wireless-specific failure and escalate to the appropriate support group.

15. Review Error Messages Before Escalation

Capture exact communication errors before sending the device for repair.

Action:

Record any displayed messages related to:

Take a photo of the error screen if allowed by facility policy and no patient information is exposed.

Expected outcome: The repair team, PACS team, or vendor has useful information for diagnosis.

If the Problem Persists

If patient/order issues, worklist filters, wireless coverage, PACS availability, destination selection, login status, and basic network recovery have been ruled out, the issue may involve internal network configuration, wireless hardware, software services, DICOM settings, or system-level failure.

The DRX-Revolution should be:

Knowing when to stop is proper troubleshooting. Repeated rebooting, deleting studies, changing DICOM settings without authorization, or performing unnecessary repeat exposures can create patient safety and documentation risks.

Clinical Use Tip

Do not troubleshoot PACS transfer problems on an active patient or during an urgent portable exam. Complete patient care first, protect locally stored images, and move to another mobile X-ray system if imaging workflow is affected. Never repeat an exposure just because an image has not appeared in PACS until local image status and transfer status have been verified.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Technologist reported that the Carestream DRX-Revolution completed portable chest images but the study would not send to PACS and the worklist would not refresh."

Cause

What was observed during troubleshooting.

Example:
"Found the unit connected to weak wireless coverage in a patient care area, with images pending in the send queue and worklist query timing out."

Resolution

What action was taken.

Example:
"Moved unit to known-good wireless area, confirmed network reconnection, manually resent pending images successfully, refreshed worklist, and returned unit to service."

Helpful Details to Include (If Known)

Final Thought

PACS and worklist failures on a mobile X-ray system should be approached logically. Start with patient safety, protect completed images, verify the order and workflow, then check wireless connectivity, destination selection, and external PACS/RIS availability. Escalation is appropriate once external causes are ruled out, especially when images cannot be transmitted reliably. Clear CCR documentation helps separate device failure from network, workflow, and PACS issues.

That is successful troubleshooting.