Cepheid GeneXpert Infinity

Clot, Bubble, or Sample-Volume Error Repeats

On this page

Asset Type

Molecular Diagnostic System

Manufacturer

Cepheid

Model

GeneXpert Infinity

What This Guide Helps With

Troubleshoot repeated sample-quality or sample-volume errors by separating specimen preparation, cartridge, loading, and analyzer-related causes.

Step-by-Step Troubleshooting

1. Protect the Patient Specimen

Stop repeated attempts if continued testing could consume or compromise a limited patient specimen. Coordinate with laboratory staff regarding recollection, alternate testing, or use of another validated analyzer.

Expected outcome: The specimen is preserved as much as possible and patient testing remains supported.

2. Record the Exact Error

Capture the complete displayed error and identify the assay, cartridge, specimen, and module involved. Determine whether the message specifically identifies sample volume or whether staff are describing a suspected clot or bubble.

Expected outcome: The actual system-reported condition is documented accurately.

3. Have Laboratory Staff Verify Specimen Quality

Clinical laboratory personnel should inspect and assess specimen suitability according to their validated procedure, including any concerns involving inadequate volume, clots, bubbles, viscosity, or preparation.

Clinical Engineering should not make independent clinical acceptance decisions for specimens.

Expected outcome: Specimen-related causes are identified or reasonably ruled out by laboratory staff.

4. Verify Required Sample Preparation

Confirm with laboratory staff that the assay-specific specimen preparation process was followed and that no preparation step was missed or altered.

Expected outcome: The specimen was prepared according to the validated assay workflow.

5. Inspect the Cartridge

Inspect the cartridge externally for damage, leakage, contamination, deformation, or improper identification. Verify it is the correct cartridge for the selected assay.

Do not open or manipulate the internal contents of the cartridge.

Expected outcome: The cartridge is externally intact and appropriate for the test.

6. Verify Loading and Position

Confirm the cartridge is fully seated and correctly positioned in the intended loading workflow. Inspect accessible areas for debris or obstructions.

Expected outcome: Loading and positioning are correct.

7. Determine Whether the Error Follows the Specimen

Where laboratory procedures permit, compare results with another properly prepared specimen or approved verification material.

Expected outcome: If the error follows the original specimen, a sample-related cause becomes more likely.

8. Determine Whether the Error Follows a Cartridge or Module

Review whether repeated errors are associated with one cartridge lot, one individual cartridge, or one analyzer module.

Expected outcome: A repeatable consumable or module pattern is identified or ruled out.

9. Verify System Readiness

Confirm the system is fully initialized and free of broader module, automation, motion, temperature, or communication errors.

Expected outcome: No other equipment condition explains the repeated processing error.

10. Verify Successful Processing

After an external cause is corrected, have laboratory staff perform the appropriate approved verification.

Expected outcome: The sample-processing error does not recur and laboratory acceptance requirements are satisfied. Troubleshooting can stop.

11. Escalate Persistent Repeated Errors

If acceptable specimens, proper preparation, intact cartridges, correct loading, and system readiness have been confirmed but the error repeats, stop troubleshooting at the external level.

Expected outcome: The affected module or system is prevented from producing potentially unreliable results until serviced.

If the Problem Persists

Common specimen, cartridge, preparation, positioning, and loading causes have been ruled out. The remaining problem may involve an analyzer module, internal cartridge-interface mechanism, sensor, thermal or mechanical subsystem, or other service-level condition.

The affected GeneXpert Infinity module or system should be:

Laboratory acceptance testing must be completed before returning the affected pathway to patient testing.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Repeated runs can consume irreplaceable specimen volume; coordinate with the laboratory before attempting another test on a limited sample.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory staff reported repeated sample-volume errors when processing a patient specimen on the GeneXpert Infinity."

Cause

What was observed during troubleshooting.

Example:
"Laboratory review found the original specimen volume inadequate, while Clinical Engineering verified normal system operation with approved verification material."

Resolution

What action was taken.

Example:
"Clinical Engineering documented normal analyzer operation, laboratory staff redirected the inadequate specimen according to policy, and the analyzer remained available for clinical use."

Helpful Details to Include (If Known)

Final Thought

Repeated sample errors require cooperation between Clinical Engineering and the laboratory. Protect the specimen, separate sample and consumable causes from equipment causes, and escalate only after the external factors are properly evaluated.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide