Cepheid GeneXpert Infinity

Result Is Delayed, Missing, or Inconsistent

On this page

Asset Type

Molecular Diagnostic System

Manufacturer

Cepheid

Model

GeneXpert Infinity

What This Guide Helps With

Troubleshoot delayed, missing, or inconsistent results by separating assay completion, specimen, cartridge, module, software, communication, and result-routing causes.

Step-by-Step Troubleshooting

1. Protect Clinical Decision-Making

Notify laboratory personnel that the affected result pathway is under evaluation. Do not allow clinicians to rely on an incomplete, inconsistent, or unverified result.

Urgent specimens should be redirected according to laboratory downtime or alternate-testing procedures when necessary.

Expected outcome: Clinical decisions are not based on questionable or unavailable results.

2. Confirm What Happened to the Result

Determine whether the assay failed to complete, completed without displaying a result, displayed locally but did not reach the LIS, produced an unexpected result pattern, or was significantly delayed.

Record the specimen, assay, cartridge, module, timestamps, and exact messages involved.

Expected outcome: The problem is narrowed to test processing, result generation, local display, or downstream transmission.

3. Verify Patient and Specimen Identification

Confirm with laboratory staff that the correct specimen, patient identifier, barcode, test order, and assay were associated with the run.

Do not correct patient identification outside established laboratory procedures.

Expected outcome: The result is associated with the intended specimen and order.

4. Review the Test Status

Determine whether the test completed normally, stopped with an error, remained in progress, or was interrupted. Check for power, module, automation, or workstation events occurring during the run.

Expected outcome: The system state explains whether the issue occurred before or after result generation.

5. Inspect Cartridge and Sample Factors

With laboratory staff, review whether the cartridge was appropriate and intact and whether sample preparation met the validated workflow.

Look for a pattern involving a particular cartridge, lot, specimen type, or assay.

Expected outcome: Obvious sample or consumable causes are identified or ruled out.

6. Determine Whether the Problem Follows a Module

Review whether delays or inconsistent results repeatedly occur on one GeneXpert module while comparable testing on other available modules is normal.

Expected outcome: A module-specific pattern is identified or ruled out.

7. Check Workstation and Application Status

Confirm the workstation is responsive and the GeneXpert application is functioning normally. Look for frozen screens, incomplete synchronization, storage warnings, or communication messages visible to the user.

Do not delete data or alter databases during troubleshooting.

Expected outcome: The local system can display and manage completed testing normally.

8. Check LIS and Network Routing

If a result exists locally but is missing downstream, inspect network connections and coordinate with laboratory IT or interface support to verify LIS and interface availability.

Expected outcome: The missing result is localized to the analyzer or downstream communication path.

9. Compare With Other Recent Tests

Review whether other specimens processed before and after the affected test completed and transmitted normally. Avoid drawing conclusions from a single unusual patient result without laboratory review.

Expected outcome: The issue is identified as isolated or part of a repeatable system pattern.

10. Perform Approved Functional Verification

After correcting an external issue, have laboratory staff perform appropriate approved verification using suitable material and confirm both local completion and downstream result transmission when applicable.

Expected outcome: Results complete consistently, display properly, and reach the intended system. If verified, troubleshooting can stop.

11. Escalate Unresolved Result Problems

If identification, specimen handling, cartridge factors, module behavior, workstation status, network communication, and result routing are all verified but results remain delayed, missing, or inconsistent, stop external troubleshooting.

Expected outcome: An unreliable result pathway is removed from clinical use pending qualified evaluation.

If the Problem Persists

Common external specimen, cartridge, workflow, workstation, communication, and infrastructure causes have been ruled out. Remaining causes may involve an analyzer module, processing subsystem, software, database, interface configuration, or another service-level condition.

The affected GeneXpert Infinity system or module should be:

Following repair, verify test completion, result consistency, and the entire result-routing pathway. Laboratory quality requirements must be satisfied before patient testing resumes.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

When a result is missing, always determine whether the test failed to generate a result or whether a valid result exists but failed to reach the LIS.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory staff reported that completed GeneXpert Infinity results were visible locally but missing from the LIS."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found loss of network communication between the analyzer workstation and the laboratory network while local testing remained complete."

Resolution

What action was taken.

Example:
"Clinical Engineering restored the external network connection and coordinated verification that a completed test result successfully transmitted through the interface to the LIS."

Helpful Details to Include (If Known)

Final Thought

Separate result generation from result transmission before assuming analyzer failure. Protect clinical decision-making, verify each step of the testing and communication pathway, escalate unresolved faults, and document the confirmed cause and final verification.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide