On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoots persistent alarms or unreliable alarm sound caused by unresolved conditions, setup problems, settings, obstructions, or interface failure.
Step-by-Step Troubleshooting
1. Protect the Patient and Respond to the Alarm Condition
Treat an active alarm as valid until the underlying condition has been identified. If the device cannot provide reliable alarm annunciation during a procedure, substitute a verified system.
Do not troubleshoot alarm audio while relying on that alarm for patient or procedural safety.
Expected outcome: Clinical care does not depend on an alarm system known to be unreliable.
2. Capture the Exact Alarm
Record:
- Exact displayed message.
- Visual indicator behavior.
- Whether audio is present, weak, intermittent, or absent.
- When the alarm occurs.
- What function the device was performing at the time.
Do not clear the alarm before gathering useful information when doing so is safe.
Expected outcome: The complaint is tied to a specific alarm state or audio problem.
3. Resolve the External Condition Before Judging the Alarm
Inspect the relevant external setup indicated by the alarm. Depending on the condition, check appropriate:
- Tubing.
- Connections.
- Consumables.
- Accessories.
- Procedure interfaces.
- Power source.
- Airflow around the unit.
Expected outcome: Any genuine external alarm condition is corrected before evaluating whether the alarm itself is malfunctioning.
4. Verify Normal Accessory and Tubing Setup
Check for kinks, disconnections, occlusions, loose fittings, misinstalled consumables, or incompatible accessories.
Expected outcome: The system setup is correct and does not continue to generate a legitimate alarm.
5. Verify User-Accessible Alarm Settings
Review normal operator-accessible audio or alarm settings if applicable. Confirm they have not been inadvertently changed.
Do not disable required alarms or enter restricted service menus.
Expected outcome: Alarm settings are appropriate for clinical use.
6. Check for Obstructed Sound
Inspect accessible speaker openings for tape, covers, drapes, packaging, heavy contamination, or placement against another object that could attenuate alarm sound.
Do not insert tools into speaker openings.
Expected outcome: The audible alarm path is unobstructed.
7. Perform a Controlled Restart When Appropriate
If the underlying alarm condition has been resolved but the alarm remains latched unexpectedly, remove the device from patient use and perform one normal restart.
Expected outcome: The device starts normally without recurrence of an unsupported alarm.
8. Verify Visual and Audible Alarm Operation
Using approved procedures, verify that applicable alarms are both visible and audible and that the system responds appropriately when the alarm condition is cleared.
Do not intentionally create unsafe conditions solely to test an alarm.
Expected outcome: Alarm annunciation is clear, appropriate, and repeatable.
9. Verify Operation After Correction
Operate the system in an approved test configuration and confirm the original alarm does not recur without cause.
Expected outcome: Alarm behavior remains normal and audio is clearly perceptible. Troubleshooting can stop.
10. Escalate Persistent or Missing Alarm Behavior
If alarm audio remains weak or absent, or an alarm persists despite correction of the documented external condition, remove the device from service.
Expected outcome: Unreliable alarm behavior is escalated rather than accepted for clinical operation.
If the Problem Persists
Common external causes have been ruled out. Remaining possibilities include an internal audio component, alarm circuitry, sensing problem, software, configuration, or other service-level fault.
The device should be:
- Removed from service.
- Labeled Out of Service.
- Sent for repair or bench evaluation.
- Evaluated using appropriate CONMED documentation and approved test equipment.
- Repaired or configured only by qualified personnel.
Before return to service, verify both visual and audible alarm performance along with the affected operating function.
Knowing when to stop external troubleshooting and escalate is proper troubleshooting.
Clinical Use Tip
An alarm that cannot be heard reliably should be treated as a safety failure even if the rest of the device appears to function normally.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"OR staff reported a persistent AirSeal iFS alarm during room setup."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found a kinked external tubing segment producing the alarm condition and verified normal alarm audio."
Resolution
What action was taken.
Example:
"Corrected the tubing routing, confirmed the alarm cleared, verified normal audible and visual alarm operation, and returned the system to service."
Helpful Details to Include (If Known)
- Exact alarm message.
- Visual alarm behavior.
- Audible alarm behavior.
- Operating condition when the alarm occurred.
- Tubing and accessory condition.
- Consumables inspected or replaced.
- Alarm settings observed.
- Speaker opening condition.
- Restart result.
- Final alarm verification.
Final Thought
Persistent alarms should be investigated as real conditions first, while missing or weak audio should be treated as a safety concern. Verify the setup, correct the cause, confirm alarm performance, and escalate whenever annunciation remains unreliable.
That is successful troubleshooting.