CONMED AirSeal iFS

Self-Test or Calibration Will Not Complete

On this page

Asset Type

Surgical Smoke Evacuator

Manufacturer

CONMED

Model

AirSeal iFS

What This Guide Helps With

Troubleshoots incomplete self-tests or calibration routines by checking setup, approved accessories, connections, environmental conditions, and test prerequisites.

Step-by-Step Troubleshooting

1. Remove the System From Clinical Dependence

Do not perform troubleshooting, calibration, or repeated self-tests while a patient depends on the AirSeal iFS.

If a required self-test or calibration cannot complete before use, substitute a verified system.

Expected outcome: The affected device is available for controlled technical evaluation without affecting patient care.

2. Record the Exact Failure

Determine:

Do not translate a generic failure into a component diagnosis.

Expected outcome: The test failure is documented accurately enough to reproduce.

3. Verify Normal Startup and Stable Power

Confirm the device completes normal startup before initiating any available approved test. Verify the power cord, receptacle, and AC stability.

Expected outcome: The test is being attempted from a normal system state with verified power.

4. Confirm Required Setup Conditions

Verify that all operator-accessible prerequisites for the test are met according to approved manufacturer documentation.

Check relevant:

Do not enter restricted service functions or change protected configuration values.

Expected outcome: The test setup matches the approved prerequisites.

5. Inspect Connected Accessories and Tubing

Look for:

Expected outcome: External components are correctly installed and free of visible defects.

6. Substitute Known-Good Approved Components

If the self-test depends on a removable accessory or disposable, repeat the approved test with a known-good compatible component when appropriate.

Expected outcome: Successful completion with a known-good component identifies the original accessory or disposable as the likely cause.

7. Check Environmental and Airflow Conditions

Confirm ventilation is not blocked and that the device is being tested in a stable indoor environment. Ensure surrounding equipment or drapes are not interfering with airflow or connections.

Expected outcome: Environmental conditions do not interfere with the test.

8. Perform One Controlled Restart and Retest

After correcting external setup issues, perform a normal restart and repeat the approved self-test or calibration once.

Avoid repeatedly clearing or restarting the system to force a pass.

Expected outcome: The test completes successfully and the system reaches the expected ready state.

9. Verify Normal Operation After a Successful Test

A successful self-test should be followed by functional verification of the relevant system functions and alarms.

Expected outcome: The test completes normally and the associated operating functions are reliable. Troubleshooting can stop.

10. Escalate Repeated Test Failure

If the test repeatedly fails with verified power, correct setup, and known-good approved external components, stop external troubleshooting.

Expected outcome: The unit is removed from service for service-level diagnosis instead of being repeatedly reset or adjusted.

If the Problem Persists

Common external causes have been ruled out. Remaining possibilities include internal sensing, pressure or flow control, calibration data, software, internal communication, or another service-level condition.

The device should be:

Any service calibration should be completed only using authorized procedures and appropriate test equipment. Perform required functional and safety verification before return to service.

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

A failed required self-test is a reason to substitute the system, not a condition to bypass because basic functions appear to operate.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Staff reported that the AirSeal iFS would not complete its pre-use system check."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found a procedure tubing connection incompletely seated and verified that no connector damage was present."

Resolution

What action was taken.

Example:
"Corrected the tubing connection, repeated the system check successfully, verified normal operation, and returned the unit to service."

Helpful Details to Include (If Known)

Final Thought

A failed self-test should be approached by confirming the setup and external test conditions before assuming calibration or internal hardware failure. If a correctly configured system repeatedly fails, remove it from service and escalate with clear documentation.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide