CONMED System 2450 / 5000 Series

Error: Err 114 Display or keyboard SPI communication failure

On this page

Asset Type

Electrosurgical Unit (ESU)

Manufacturer

CONMED

Model

System 2450 / 5000 Series

What This Guide Helps With

This guide helps troubleshoot Err 114 on a CONMED System 2450 / 5000 Series electrosurgical unit. Err 114 indicates a display or keyboard SPI communication failure, which may cause control panel problems, display issues, or unreliable user input. Because the operator may not be able to confirm settings or control the ESU correctly, the unit should not be used clinically until the issue is resolved.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Cycle System Power

Check the Power Source and Startup Conditions

Inspect the Front Panel and Display Area

Remove External Accessories and Retest

Verify Keyboard and Display Function

Inspect Accessible Cable Connections During Bench Evaluation

Isolate the Likely Display, Cable, or Controller Failure

Perform Advanced Signal Verification If Qualified

If the Problem Persists

If Err 114 continues after power cycling, power source verification, accessory removal, front panel inspection, keyboard/display checks, and accessible cable inspection, external causes have been ruled out.

A persistent Err 114 indicates a likely internal display, keyboard communication cable, Display assembly, or Controller assembly failure.

Remove the device from service, label it Out of Service, and send it for repair or qualified bench evaluation. Do not return the ESU to clinical use with unresolved display or keyboard communication faults.

Clinical Use Tip

Never troubleshoot an ESU communication error on an active patient. If the display or keyboard is unreliable, staff may not be able to confirm power settings, mode selection, or device status. Move the patient to a backup device first, then evaluate the affected ESU away from patient use.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"CONMED System 2450 / 5000 Series ESU displayed Err 114 display or keyboard SPI communication failure during startup. Staff reported display/control panel communication issue."

Cause

What was observed during troubleshooting.

Example:
"Confirmed Err 114 on startup. Power source, external accessories, front panel condition, and keyboard/display response were checked. Fault persisted, indicating likely internal display-to-controller communication issue involving cable, Display assembly, or Controller assembly."

Resolution

What action was taken.

Example:
"Removed ESU from service and labeled Out of Service. Sent for bench evaluation/repair of Display-to-Controller cable, Display assembly, or Controller assembly. Unit not returned to service pending repair and functional verification."

Helpful Details to Include (If Known)

Final Thought

Err 114 should be treated as a real communication failure until proven otherwise. Start with patient safety, confirm the error, rule out simple external causes, and then isolate the display, keyboard, cable, and controller communication path. If the fault persists, remove the ESU from service and document the complaint, observed cause, and resolution clearly.

That is successful troubleshooting.