CONMED System 2450 / 5000 Series

Error: Err 117 5.0 volt too low

On this page

Asset Type

Electrosurgical Unit (ESU)

Manufacturer

CONMED

Model

System 2450 / 5000 Series

What This Guide Helps With

This guide helps troubleshoot Err 117 on a CONMED System 2450 / 5000 Series electrosurgical unit when the generator detects a low 5.0 V supply condition. This fault can affect controller operation, monitoring circuits, display behavior, activation response, and startup safety checks.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check the AC Power Source First

Inspect the Power Cord and External Condition

Power Cycle and Recheck the Error

Remove Accessories and Retest

Check Basic Front Panel Behavior

Escalate to Bench Evaluation

If the Problem Persists

If Err 117 continues after verifying the AC outlet, power cord, accessories, external condition, and power cycling, the common external causes have been ruled out.

A persistent Err 117 strongly indicates an internal low-voltage power supply problem or Controller assembly voltage monitoring fault.

The device should be removed from service, labeled Out of Service, and sent for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot this error on an active patient. Move the patient or procedure to a backup ESU first. A low 5.0 V logic supply fault can affect system control, monitoring, display function, and activation reliability.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"CONMED System 2450 / 5000 Series ESU displayed Err 117, 5.0 volt too low, during startup. Unit removed from service due to power supply-related system fault."

Cause

What was observed during troubleshooting.

Example:
"Verified Err 117 was repeatable. External AC outlet, power cord, accessories, and basic setup were checked. Fault indicates low 5.0 V supply condition or Controller voltage monitoring issue."

Resolution

What action was taken.

Example:
"Removed ESU from clinical use and labeled Out of Service. Unit requires bench evaluation of low-voltage power supply output and Controller ADC reference voltages. Repair to include power supply replacement if output is out of limits or Controller assembly replacement if supply voltage is normal and fault persists."

Helpful Details to Include (If Known)

Final Thought

Err 117 should be treated as a system power integrity fault, not a normal user setup issue. Start with patient safety, confirm the error, rule out simple external causes, then escalate appropriately when the low-voltage supply or Controller assembly is suspected. Clear documentation helps show that the fault was handled logically and safely.

That is successful troubleshooting.