CONMED System 2450 / 5000 Series

Error: Err 136 A.R.M.™ circuit dropped below 0.4 V.

On this page

Asset Type

Electrosurgical Unit (ESU)

Manufacturer

CONMED

Model

System 2450 / 5000 Series

What This Guide Helps With

This guide helps troubleshoot Err 136 on the CONMED System 2450 / 5000 Series Electrosurgical Unit when the A.R.M.™ circuit voltage drops below 0.4 V. This condition may indicate a return electrode monitoring signal issue, output assembly problem, signal path failure, or controller sensing failure.

Because A.R.M.™ monitoring is related to patient return electrode safety, this error should be treated as a patient safety concern until the device has been evaluated and verified safe for use.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Understand What the Error Means

Check Power Source and Basic Setup

Check External Accessories and Return Electrode Setup

Inspect the Output Connector Area

Check for Intermittent Behavior

Isolate External Setup From Device Failure

Bench Evaluate the A.R.M.™ Signal Path

Determine Corrective Action

Stop When Resolved

If the Problem Persists

If Err 136 continues after power, accessories, return electrode setup, connector condition, and other external causes have been ruled out, the condition strongly suggests an internal A.R.M.™ circuit, Output assembly, signal path, or Controller assembly failure.

The device should be removed from service, labeled Out of Service, and sent for repair or advanced bench evaluation.

Do not keep cycling power or returning the ESU to clinical use with Err 136 present.

Clinical Use Tip

Never troubleshoot an ESU with an active A.R.M.™ circuit error on a patient. Move the patient or procedure to a known-good backup ESU first, then evaluate the affected generator on the bench. A return electrode monitoring-related fault should always be treated as a patient safety concern until proven otherwise.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"CONMED System 2450 / 5000 Series ESU displayed Err 136 A.R.M.™ circuit voltage low error and was removed from service."

Cause

What was observed during troubleshooting.

Example:
"Err 136 confirmed during bench evaluation. External power, accessories, return electrode setup, and connector condition were checked. Fault persisted, indicating likely internal A.R.M.™ signal path, Output assembly, or Controller assembly failure."

Resolution

What action was taken.

Example:
"Removed unit from service and labeled Out of Service. Verified VARM signal path per service procedure. Unit requires Output assembly or Controller assembly repair based on VARM findings, followed by functional testing, ESU performance verification, and safety testing before return to service."

Helpful Details to Include (If Known)

Final Thought

Err 136 should be approached as a safety-related monitoring circuit fault, not just a reset-and-return issue. Start with patient safety and simple external checks, then logically isolate the return electrode setup, connector condition, A.R.M.™ signal path, Output assembly, and Controller assembly. Clear CCR documentation helps the next technician understand what was ruled out and why the device was escalated or returned to service.

That is successful troubleshooting.