CONMED System 2450 / 5000 Series

Error: Err 357 Monitor Sensed Power While the Dial Power Setting Was 0 W

On this page

Asset Type

Electrosurgical Unit (ESU)

Manufacturer

CONMED

Model

System 2450 / 5000 Series

What This Guide Helps With

This guide helps troubleshoot Err 357 on a CONMED System 2450 / 5000 Series electrosurgical unit when the monitor senses output power while the dial power setting is 0 W. This condition may indicate an accessory issue, setting problem, calibration problem, false power sensing issue, or internal RF output/control failure.

Because this error involves possible sensed RF output when output should be zero, it should be treated as a patient safety concern until the ESU is removed from use, evaluated, and verified safe before return to service.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check the Power Setting and Front Panel Controls

Remove External Accessories

Inspect External Connections and Physical Condition

Power Cycle and Recheck

Perform Bench Testing Only If Qualified and Equipped

Recalibrate if Appropriate

Evaluate Likely Internal Failure Areas

If the Problem Persists

If Err 357 remains after external causes, accessories, power setting issues, outlet concerns, and common setup problems have been ruled out, remove the ESU from service.

Label the device Out of Service.

Send it for repair or advanced bench evaluation.

Do not continue repeated clinical testing with a generator that reports sensed power while the commanded power setting is 0 W. Persistent Err 357 strongly suggests an internal output control, sensing, HVPS, RF amplifier, RF output, or controller-related failure.

Clinical Use Tip

Never troubleshoot an ESU with an unresolved power-sensing fault on an active patient. Move the patient or procedure to a backup ESU first, then evaluate the affected generator in a safe bench or non-patient-care setting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"User reported CONMED System 2450 / 5000 Series ESU displayed Err 357. Error indicates monitor sensed output power while dial power setting was 0 W. Unit removed from patient use for evaluation."

Cause

What was observed during troubleshooting.

Example:
"Confirmed Err 357 during evaluation. Power setting, accessories, external connections, and power source were checked. Fault persisted with accessories removed, indicating likely internal output sensing, RF control, HVPS, RF amplifier, RF output, or controller-related failure."

Resolution

What action was taken.

Example:
"Removed ESU from service and labeled Out of Service. Unit sent for repair or advanced bench evaluation. Safety and performance testing required before return to service."

Helpful Details to Include (If Known)

Final Thought

Err 357 should be treated seriously because the ESU is reporting sensed power when output should be zero. Good troubleshooting starts with patient safety, confirms the error, rules out settings and accessories, then moves to controlled bench testing, calibration, and repair escalation when needed. Clear CCR documentation protects the patient, the technician, and the equipment history.

That is successful troubleshooting.