CONMED System 2450 / 5000 Series

Error: Err 373 Damping circuit failed POST — damping not detected

On this page

Asset Type

Electrosurgical Unit (ESU)

Manufacturer

CONMED

Model

System 2450 / 5000 Series

What This Guide Helps With

This guide helps troubleshoot Err 373 on a CONMED System 2450 / 5000 Series electrosurgical unit when the generator fails POST because the damping circuit is not detected. This condition can prevent safe RF output operation and should be treated as a serious generator fault until the device is evaluated and tested.

Step-by-Step Troubleshooting

Ensure Patient Safety First

The patient should not be exposed to an ESU with an unresolved RF control or damping fault. A damping circuit fault may affect RF output control, especially during coagulation modes.

Verify the Reported Error and Behavior

This confirms the issue is a true system-reported damping circuit fault and not a user setup issue, accessory problem, or unrelated activation complaint.

Check Power Source and Basic Setup

Stable input power should be confirmed before assuming an internal control or RF circuit failure.

Inspect the Unit Externally

No external damage or accessory issue should explain the POST failure before the unit is escalated for bench-level evaluation.

Power Cycle and Recheck

A temporary logic fault may clear, but a repeatable POST failure strongly suggests the unit is detecting a real internal damping circuit or control problem.

Isolate From User Setup

The damping circuit is internal to the RF output/control path, but external setup should still be ruled out before repair escalation.

Configure for Bench Evaluation or Test Mode

Err 373 involves RF drive and damping timing. Troubleshooting should not be performed casually in a clinical area.

Verify Damping-Related Operation During Coag Modes

Err 373 means the system expected to detect damping behavior during POST but did not. This points toward the RF Amplifier damping circuitry or Controller assembly.

Inspect Internal Connections Only at an Appropriate Service Level

A missing damping signal may be caused by an actual circuit failure, but connector issues should be ruled out before replacing assemblies.

Determine Whether the Fault Is Device-Related

Err 373 is not a normal user-correctable alarm. Persistent failure indicates the ESU cannot verify a required RF damping function.

Stop When Resolved

The ESU should pass POST, function normally, and be confirmed safe before return to clinical use.

If the Problem Persists

If Err 373 continues after external checks, power cycling, accessory isolation, and appropriate bench evaluation, external and common causes have been ruled out.

Persistent Err 373 strongly indicates an internal RF damping circuit, RF Amplifier assembly, or Controller assembly failure.

Knowing when to stop is proper troubleshooting. An ESU with an unresolved POST damping circuit failure should not be placed back into patient use.

Clinical Use Tip

Never troubleshoot an ESU with Err 373 on an active patient. Move the patient or procedure to a backup generator first, then remove the affected ESU from clinical use. An unresolved RF output control fault can create unpredictable surgical performance and should be treated as a patient safety issue.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"CONMED System 2450 / 5000 Series ESU reported Err 373 during startup. Unit failed POST with damping circuit not detected and was unavailable for clinical use."

Cause

What was observed during troubleshooting.

Example:
"Err 373 confirmed during power-on testing. External accessories disconnected and power source checked. Error persisted, indicating likely internal damping circuit, RF Amplifier, or Controller assembly failure."

Resolution

What action was taken.

Example:
"Removed ESU from service and labeled Out of Service. Sent for bench repair/evaluation. RF Amplifier damping circuitry and Controller assembly to be evaluated before return to service."

Helpful Details to Include (If Known)

Final Thought

Err 373 should be treated as a serious ESU POST failure because the generator is reporting that damping was not detected. Start with patient safety, confirm the exact error, rule out simple external causes, then move logically toward internal RF/control troubleshooting. If the fault persists, remove the device from service and escalate for repair. Clear CCR documentation protects the patient, the technician, and the facility.

That is successful troubleshooting.