CONMED System 2450 / 5000 Series

Error: Err 387 POST HVDC to 140V Failed — Supply Too High

On this page

Asset Type

Electrosurgical Unit (ESU)

Manufacturer

CONMED

Model

System 2450 / 5000 Series

What This Guide Helps With

This guide helps troubleshoot CONMED System 2450 / 5000 Series Error Err 387, where POST detects that the HVDC supply intended for the 140V startup check is too high. This can prevent the electrosurgical generator from completing startup self-test, lock out activation, and indicate an internal high-voltage power supply regulation or controller feedback problem.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Power Cycle the System

Check the Power Source

Remove External Accessories

Inspect External Condition

Confirm the Fault Is Internal to the Generator

Bench Evaluation Only

Determine the Repair Path

If the Problem Persists

If Err 387 continues after power cycling, removing external accessories, checking the power cord, and testing from a known-good AC outlet, the common external causes have been ruled out.

A persistent Err 387 strongly indicates an internal generator issue involving HVPS regulation, controller command or feedback, or related internal cabling.

The device should be removed from service, labeled Out of Service, and sent for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot an ESU with Err 387 on an active patient or during an active procedure. Move the patient or procedure to a backup generator first, then evaluate the failed unit away from clinical use.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"CONMED System 2450 / 5000 Series ESU displayed Err 387 during startup and failed POST. Unit was not available for clinical use."

Cause

What was observed during troubleshooting.

Example:
"Err 387 repeated during startup after power cycle and known-good outlet check. Accessories were removed and fault remained. External causes ruled out. Fault indicates likely internal HVPS regulation, controller command, or feedback issue."

Resolution

What action was taken.

Example:
"Removed ESU from service and labeled Out of Service. Sent for bench evaluation. Test mode verification of HVPS control signals and HVPS operation required. Repair path includes Controller assembly replacement or HVPS repair/replacement based on bench findings."

Helpful Details to Include (If Known)

Final Thought

Err 387 should be treated as a serious ESU startup fault because the generator is reporting that the 140V HVDC supply is too high during POST. Start with patient safety, confirm the exact error, rule out simple external causes, then escalate when the fault remains. Clear documentation shows that Clinical Engineering isolated the problem logically and removed the device from service when internal failure was likely.

That is successful troubleshooting.