Datascope System 98 / 98XT

Helium Cylinder or Pneumatic Leak Fault

On this page

Asset Type

Intra-Aortic Balloon Pump

Manufacturer

Datascope

Model

System 98 / 98XT

What This Guide Helps With

Troubleshooting helium supply, cylinder seating, tubing, connector, or external pneumatic leak issues on Datascope System 98 / 98XT IABPs.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm whether the Datascope System 98 / 98XT is connected to an active patient.

If the IABP is supporting a patient and there is a helium or pneumatic leak fault, do not continue extended troubleshooting at bedside. Notify clinical staff immediately and have a backup IABP available or transfer support per facility policy.

Expected: Patient support is maintained without relying on a device with an unresolved pneumatic fault.

Confirm the Reported Fault

Verify the exact alarm, fault message, or behavior reported by clinical staff.

Look for symptoms such as:

Expected: The helium or pneumatic fault is confirmed before parts are swapped.

Check Helium Cylinder Level

Inspect the installed helium cylinder and verify it is not empty or near empty.

Replace the cylinder with a known full, approved helium cylinder if the pressure appears low or uncertain.

Expected: If the issue was caused by a depleted cylinder, the fault clears after cylinder replacement and normal pneumatic operation returns. If resolved, stop troubleshooting.

Verify Cylinder Is Fully Seated

Remove and reinstall the helium cylinder following normal external handling procedures.

Confirm the cylinder is seated squarely, tightened correctly, and not cross-threaded or partially engaged.

Expected: A properly seated cylinder should seal without leakage and allow the IABP to pressurize normally. If resolved, stop troubleshooting.

Inspect the Cylinder Seal and Connection Area

Check the cylinder connection area for damaged threads, missing or damaged sealing surfaces, debris, or signs of gas leakage.

Do not use damaged cylinders or damaged connection hardware.

Expected: The cylinder connection should be clean, intact, and free of visible damage. Damaged cylinders should be removed from use.

Listen for External Leakage

With the device powered and attempting normal pneumatic operation, listen near the helium cylinder connection, rear pneumatic fittings, catheter connection area, and accessible external tubing.

Do not open the unit or perform internal pneumatic disassembly.

Expected: No continuous hissing should be heard. A clear external leak location may identify a cylinder, fitting, tubing, or connection problem.

Check the IAB Catheter Pneumatic Connection

Inspect the balloon catheter tubing connection to the pump.

Confirm the catheter extension tubing is fully connected, not kinked, cracked, loose, or contaminated.

Expected: The catheter pneumatic path should be secure and unobstructed. If reseating the connection restores operation, document the finding and return the device only after appropriate functional verification.

Check for Kinks or Damage in External Tubing

Inspect accessible external pneumatic tubing for sharp bends, pinching, stretching, cracks, or loose fittings.

Pay close attention to tubing routed around bed rails, transport hardware, or the IABP cart.

Expected: Tubing should be intact and routed without restriction. If corrected and the fault clears, stop troubleshooting after confirming normal operation.

Swap Known-Good External Accessories When Safe

If available and appropriate, use a known-good helium cylinder and approved external tubing or connection accessories.

Do not swap patient-connected components unless clinical staff confirms it is safe and clinically appropriate.

Expected: If the fault follows an accessory, remove that accessory from service. If the fault stays with the IABP, suspect an internal pneumatic issue.

Verify Device Startup and Self-Test Behavior

Power cycle the device only when it is not supporting a patient or when clinical staff has confirmed it is safe.

Observe whether the helium or pneumatic fault appears immediately at startup or only during pumping.

Expected: A fault present immediately at startup with a known full cylinder and secure connections suggests an internal pneumatic or sensing problem.

Check for Signs of Regulator or Internal Pneumatic Failure

If the cylinder is full, properly seated, and no external leak is found, note whether the pump still fails to pressurize, vents helium rapidly, or repeatedly alarms during assist.

Expected: Persistent failure after external checks points toward an internal pneumatic leak, regulator issue, valve problem, pressure sensor issue, or internal tubing failure.

Remove From Service if the Fault Persists

If the fault remains after checking cylinder level, cylinder seating, visible pneumatic connections, catheter tubing, and external accessories, remove the IABP from service.

Expected: The device is not returned to clinical use until evaluated and repaired by qualified service personnel.

If the Problem Persists

At this point, the common external causes have been ruled out. The issue is likely internal to the pneumatic system, regulator assembly, valve pathway, pressure sensing circuit, or internal tubing.

The Datascope System 98 / 98XT should be:

Knowing when to stop is proper troubleshooting. An IABP with an unresolved helium or pneumatic leak fault should not be used for patient support.

Clinical Use Tip

Do not troubleshoot a helium or pneumatic leak fault on an active patient unless immediate safety requires a quick external check. Move support to another verified IABP or have clinical staff manage the patient first. Pneumatic reliability is critical because balloon inflation and deflation timing depend on stable helium delivery.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported the Datascope System 98 / 98XT displayed a helium cylinder or pneumatic leak fault and would not maintain balloon assist pressure."

Cause

What was observed during troubleshooting.

Example:
"A full helium cylinder was installed and external cylinder seating, catheter tubing, and visible pneumatic connections were checked, but the fault remained with the pump."

Resolution

What action was taken.

Example:
"IABP was removed from service, labeled Out of Service, and sent for bench evaluation due to suspected internal pneumatic leak or regulator failure."

Helpful Details to Include (If Known)

Final Thought

Helium and pneumatic faults on an intra-aortic balloon pump must be handled carefully because they can directly affect balloon inflation, deflation, and patient support. Start with the cylinder, seating, visible connections, and tubing, but stop before internal pneumatic repair unless properly authorized and equipped. Clear CCR documentation helps show that patient safety was protected and that escalation was appropriate.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide