On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting helium supply, cylinder seating, tubing, connector, or external pneumatic leak issues on Datascope System 98 / 98XT IABPs.
Step-by-Step Troubleshooting
Ensure Patient Safety First
Confirm whether the Datascope System 98 / 98XT is connected to an active patient.
If the IABP is supporting a patient and there is a helium or pneumatic leak fault, do not continue extended troubleshooting at bedside. Notify clinical staff immediately and have a backup IABP available or transfer support per facility policy.
Expected: Patient support is maintained without relying on a device with an unresolved pneumatic fault.
Confirm the Reported Fault
Verify the exact alarm, fault message, or behavior reported by clinical staff.
Look for symptoms such as:
- Helium cylinder empty or low indication
- Unable to build pneumatic pressure
- Repeated helium loss alarm
- Balloon not inflating properly
- Pump cycling but no effective assist
- Audible hiss from cylinder, regulator area, hose, or rear panel
Expected: The helium or pneumatic fault is confirmed before parts are swapped.
Check Helium Cylinder Level
Inspect the installed helium cylinder and verify it is not empty or near empty.
Replace the cylinder with a known full, approved helium cylinder if the pressure appears low or uncertain.
Expected: If the issue was caused by a depleted cylinder, the fault clears after cylinder replacement and normal pneumatic operation returns. If resolved, stop troubleshooting.
Verify Cylinder Is Fully Seated
Remove and reinstall the helium cylinder following normal external handling procedures.
Confirm the cylinder is seated squarely, tightened correctly, and not cross-threaded or partially engaged.
Expected: A properly seated cylinder should seal without leakage and allow the IABP to pressurize normally. If resolved, stop troubleshooting.
Inspect the Cylinder Seal and Connection Area
Check the cylinder connection area for damaged threads, missing or damaged sealing surfaces, debris, or signs of gas leakage.
Do not use damaged cylinders or damaged connection hardware.
Expected: The cylinder connection should be clean, intact, and free of visible damage. Damaged cylinders should be removed from use.
Listen for External Leakage
With the device powered and attempting normal pneumatic operation, listen near the helium cylinder connection, rear pneumatic fittings, catheter connection area, and accessible external tubing.
Do not open the unit or perform internal pneumatic disassembly.
Expected: No continuous hissing should be heard. A clear external leak location may identify a cylinder, fitting, tubing, or connection problem.
Check the IAB Catheter Pneumatic Connection
Inspect the balloon catheter tubing connection to the pump.
Confirm the catheter extension tubing is fully connected, not kinked, cracked, loose, or contaminated.
Expected: The catheter pneumatic path should be secure and unobstructed. If reseating the connection restores operation, document the finding and return the device only after appropriate functional verification.
Check for Kinks or Damage in External Tubing
Inspect accessible external pneumatic tubing for sharp bends, pinching, stretching, cracks, or loose fittings.
Pay close attention to tubing routed around bed rails, transport hardware, or the IABP cart.
Expected: Tubing should be intact and routed without restriction. If corrected and the fault clears, stop troubleshooting after confirming normal operation.
Swap Known-Good External Accessories When Safe
If available and appropriate, use a known-good helium cylinder and approved external tubing or connection accessories.
Do not swap patient-connected components unless clinical staff confirms it is safe and clinically appropriate.
Expected: If the fault follows an accessory, remove that accessory from service. If the fault stays with the IABP, suspect an internal pneumatic issue.
Verify Device Startup and Self-Test Behavior
Power cycle the device only when it is not supporting a patient or when clinical staff has confirmed it is safe.
Observe whether the helium or pneumatic fault appears immediately at startup or only during pumping.
Expected: A fault present immediately at startup with a known full cylinder and secure connections suggests an internal pneumatic or sensing problem.
Check for Signs of Regulator or Internal Pneumatic Failure
If the cylinder is full, properly seated, and no external leak is found, note whether the pump still fails to pressurize, vents helium rapidly, or repeatedly alarms during assist.
Expected: Persistent failure after external checks points toward an internal pneumatic leak, regulator issue, valve problem, pressure sensor issue, or internal tubing failure.
Remove From Service if the Fault Persists
If the fault remains after checking cylinder level, cylinder seating, visible pneumatic connections, catheter tubing, and external accessories, remove the IABP from service.
Expected: The device is not returned to clinical use until evaluated and repaired by qualified service personnel.
If the Problem Persists
At this point, the common external causes have been ruled out. The issue is likely internal to the pneumatic system, regulator assembly, valve pathway, pressure sensing circuit, or internal tubing.
The Datascope System 98 / 98XT should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
Knowing when to stop is proper troubleshooting. An IABP with an unresolved helium or pneumatic leak fault should not be used for patient support.
Clinical Use Tip
Do not troubleshoot a helium or pneumatic leak fault on an active patient unless immediate safety requires a quick external check. Move support to another verified IABP or have clinical staff manage the patient first. Pneumatic reliability is critical because balloon inflation and deflation timing depend on stable helium delivery.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported the Datascope System 98 / 98XT displayed a helium cylinder or pneumatic leak fault and would not maintain balloon assist pressure."
Cause
What was observed during troubleshooting.
Example:
"A full helium cylinder was installed and external cylinder seating, catheter tubing, and visible pneumatic connections were checked, but the fault remained with the pump."
Resolution
What action was taken.
Example:
"IABP was removed from service, labeled Out of Service, and sent for bench evaluation due to suspected internal pneumatic leak or regulator failure."
Helpful Details to Include (If Known)
- Helium cylinder pressure or full-cylinder replacement confirmed
- Cylinder removed and reseated
- Cylinder connection inspected for damaged threads or seal issues
- Audible hiss present or not present
- Fault occurred at startup or only during pumping
- Catheter tubing checked for kinks, cracks, or loose connection
- Known-good cylinder tested
- External tubing or accessory swapped, if applicable
- Any unusual venting, cycling, or pressure behavior
- Final device status: returned to use, removed from service, or sent for repair
Final Thought
Helium and pneumatic faults on an intra-aortic balloon pump must be handled carefully because they can directly affect balloon inflation, deflation, and patient support. Start with the cylinder, seating, visible connections, and tubing, but stop before internal pneumatic repair unless properly authorized and equipped. Clear CCR documentation helps show that patient safety was protected and that escalation was appropriate.
That is successful troubleshooting.