On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting no oxygen flow, weak outlet flow, blocked tubing, flowmeter issues, humidifier restriction, or internal concentrator performance concerns.
Step-by-Step Troubleshooting
Ensure Patient Safety First
If the concentrator is connected to a patient and there is no flow, low flow, or questionable oxygen delivery, move the patient to another approved oxygen source immediately.
Expected outcome: The patient is not dependent on a concentrator with unreliable oxygen output.
Do not continue troubleshooting while the patient relies on the affected unit.
Verify the Reported Flow Problem
Turn the DeVilbiss 525 Series on and allow it to run in a controlled Biomed setting.
Check whether the issue is:
- No flow at the patient outlet
- Weak or restricted flow
- Flowmeter ball not rising
- Flowmeter ball rises but no flow reaches the cannula
- Flow drops after the unit runs for several minutes
Expected outcome: The symptom is confirmed before parts or accessories are blamed.
Check the Flowmeter Setting
Verify the flow control knob is not closed or set below the expected liter flow.
Slowly adjust the flow control and observe whether the flowmeter ball moves smoothly.
Expected outcome: The flowmeter responds to adjustment and stabilizes at the selected flow.
If the flow was simply turned down or closed, correct the setting, verify proper outlet flow, document the finding, and stop.
Inspect the Patient Tubing and Cannula
Remove the nasal cannula and oxygen tubing from the concentrator outlet.
Inspect for kinks, crushed tubing, knots, liquid obstruction, damaged connectors, or excessive tubing length.
Replace with known-good tubing and cannula.
Expected outcome: Flow returns with known-good patient tubing.
If tubing replacement restores flow, return the device only after verifying stable flow at the prescribed setting.
Check the Humidifier Bottle, If Used
If a humidifier bottle is installed, remove it and connect known-good tubing directly to the patient outlet.
Inspect the humidifier bottle for:
- Cross-threaded cap
- Blocked diffuser stem
- Overfilled water chamber
- Mineral buildup
- Cracked bottle or cap
- Incorrect connection to inlet or outlet ports
Expected outcome: Flow improves when the humidifier is removed or replaced.
If the humidifier was the restriction, replace it with a clean, compatible humidifier bottle, verify bubbling and outlet flow, document the accessory issue, and stop.
Inspect the Patient Outlet Fitting
Check the oxygen outlet barb or fitting for cracks, looseness, debris, or blocked threads.
Confirm that tubing seats securely on the outlet.
Expected outcome: The outlet is physically intact and not blocked.
If debris or a damaged outlet fitting is found, remove the unit from service for repair or outlet fitting replacement according to facility procedure.
Check for External Filter Restriction
Inspect the external cabinet filter or intake filter area for heavy dust, lint, or blockage.
Clean or replace the external filter according to facility policy.
Make sure the concentrator has adequate clearance around the air intake and exhaust areas.
Expected outcome: The unit has unrestricted airflow for compressor operation and cooling.
If cleaning a blocked external filter restores normal flow and the unit passes performance checks, document the correction and final status.
Listen for Compressor Operation
With the unit running, listen for normal compressor sound and airflow.
Note any unusual symptoms such as:
- Compressor not starting
- Loud rattling
- Weak compressor sound
- Cycling or stopping
- Excessive heat
- Burning smell
Expected outcome: The concentrator runs smoothly without abnormal noise, odor, or overheating.
If the compressor does not appear to run or sounds abnormal, remove the unit from service.
Check Flow Without Patient Accessories Attached
Run the concentrator with no tubing, cannula, or humidifier attached and observe the flowmeter.
Expected outcome: The flowmeter ball should rise and respond to adjustment.
If the flowmeter works with no accessories attached but drops when accessories are installed, the restriction is likely external to the concentrator.
Verify Flow Using a Known-Good Flow Test Method
Use an appropriate oxygen flow test device or flow analyzer at the patient outlet.
Compare measured flow to the selected flowmeter setting.
Expected outcome: Outlet flow is reasonably consistent with the selected flow setting.
If the flowmeter indicates flow but measured outlet flow is absent or significantly low, remove the concentrator from service for bench evaluation.
Check for Alarm or Indicator Behavior
Observe the front panel for alarms or indicators during operation.
Note whether alarms occur immediately, after warm-up, or only when flow is adjusted.
Expected outcome: The unit should operate without persistent alarm conditions.
If alarms accompany restricted flow, document the alarm behavior and remove the device from service.
Allow the Unit to Run Long Enough to Reproduce the Complaint
If the reported issue is intermittent, run the concentrator at a typical clinical flow setting long enough to observe whether flow drops over time.
Watch for heat buildup, flowmeter instability, abnormal sounds, or shutdown.
Expected outcome: Flow remains stable during the observation period.
If flow becomes restricted after warm-up, suspect an internal pneumatic, compressor, valve, or sieve bed performance issue and escalate for repair.
If the Problem Persists
If the DeVilbiss 525 Series still has no flow or restricted patient outlet flow after tubing, humidifier, outlet fitting, intake filter, flowmeter setting, and basic airflow checks have been ruled out, the issue is likely internal.
Possible internal causes may include compressor weakness, internal tubing restriction, flowmeter failure, valve malfunction, pressure regulation issue, sieve bed restriction, or pneumatic system leakage.
The concentrator should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
Knowing when to stop is proper troubleshooting. Do not continue using an oxygen concentrator when delivered flow cannot be verified.
Clinical Use Tip
Do not troubleshoot a low-flow or no-flow oxygen concentrator while it is supporting a patient. Place the patient on another approved oxygen source first, then evaluate the concentrator in a safe Biomed setting.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported that the DeVilbiss 525 Series oxygen concentrator had no flow from the patient outlet when connected to nasal cannula tubing."
Cause
What was observed during troubleshooting.
Example:
"Biomed found the humidifier bottle diffuser was blocked, restricting oxygen flow even though the concentrator flowmeter responded normally."
Resolution
What action was taken.
Example:
"Replaced the humidifier bottle, verified stable outlet flow with known-good tubing, checked for abnormal alarms, and returned the concentrator to service."
Helpful Details to Include (If Known)
- Patient removed from affected concentrator before troubleshooting
- Flowmeter setting checked
- Outlet flow verified with test device
- Cannula and tubing swapped
- Humidifier bottle removed or replaced
- Patient outlet inspected
- External filter condition checked
- Alarm or indicator behavior noted
- Compressor sound normal or abnormal
- Any unusual heat, smell, or vibration
- Final device status: returned to service or removed for repair
Final Thought
No-flow complaints on oxygen concentrators should be approached logically and safely. Start with the patient, then rule out tubing, humidifier, outlet, filter, and flow setting issues before suspecting internal failure. Clear CCR documentation helps show what was tested, what was found, and why the final decision was appropriate.
That is successful troubleshooting.