Drager Apollo

Alarm: APNEA FLOW

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Apollo

What This Guide Helps With

This guide helps troubleshoot the APNEA FLOW alarm on a Drager Apollo anesthesia machine. This alarm indicates that breathing or ventilation has stopped or that expected ventilation flow is not being detected. Possible causes include insufficient fresh-gas supply, a kinked tube, a hose system leak, a disconnected circuit, or a device-related ventilation or flow detection issue.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Confirm Manual Ventilation and Patient Breathing Status

Check Ventilator Mode and Settings

Check Fresh-Gas Supply and Fresh-Gas Settings

Inspect the Breathing Circuit and Hose System

Check for Kinked, Occluded, or Restricted Tubing

Check for Circuit Leak

Check Accessories and Recently Changed Components

Isolate Whether the Issue Is External or Device-Related

Stop When Resolved

If the Problem Persists

If APNEA FLOW continues after fresh-gas supply, ventilator settings, breathing circuit, tube condition, accessories, and leaks have been checked, external causes have been ruled out.

Remove the anesthesia machine from service.

Label the device Out of Service.

Send the unit for repair or bench evaluation.

Persistent APNEA FLOW may indicate an internal ventilation, flow sensing, breathing system, or control issue that requires service evaluation.

Clinical Use Tip

Never troubleshoot an APNEA FLOW alarm on an active patient if ventilation is not confirmed. Move the patient to a backup ventilation method first, then troubleshoot the anesthesia machine only after the patient is safe.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Drager Apollo anesthesia machine displayed APNEA FLOW alarm during use. Staff reported breathing/ventilation stopped or machine was not detecting ventilation flow."

Cause

What was observed during troubleshooting.

Example:
"Checked ventilator settings, fresh-gas setting, hose system, breathing circuit, patient tube routing, filter/HME, and possible leaks. Alarm appeared related to insufficient fresh-gas flow, kinked tubing, hose system leak, or unresolved ventilation/flow detection fault."

Resolution

What action was taken.

Example:
"Patient ventilation was supported by clinical staff. Fresh-gas setting, ventilator settings, breathing circuit, hose system, and tube were inspected. External issue was corrected if found. If APNEA FLOW persisted, machine was removed from service, labeled Out of Service, and referred for repair evaluation."

Helpful Details to Include (If Known)

Final Thought

APNEA FLOW is a patient safety alarm, not just a nuisance alarm. Start with ventilation support, then work logically through settings, fresh gas, tubing, hoses, leaks, and accessories. If the alarm remains after external causes are ruled out, remove the machine from service and escalate for repair. Clear documentation protects the patient, the technician, and the clinical team.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide