Drager Apollo

Alarm: APNEA PRESSURE

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Apollo

What This Guide Helps With

This guide helps troubleshoot the APNEA PRESSURE alarm on a Drager Apollo anesthesia machine. This alarm indicates that breathing or ventilation has stopped, or that airway pressure is not being detected. Possible causes include insufficient fresh-gas supply, a leak or blockage in the tube or hose system, or the patient not being connected to the breathing system.

This should be treated as a patient-critical ventilation alarm. Patient ventilation must be confirmed before troubleshooting the anesthesia machine, breathing circuit, or accessories.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Confirm Manual Ventilation and Backup Ventilation Readiness

Check Fresh-Gas Supply and Settings

Check the Patient Connection

Inspect the Breathing Circuit for Leaks

Check for Blockage or Kinking in the Tube or Hose System

Check Ventilation Mode and Settings

Isolate Accessory-Related Causes

Perform a System Check Before Returning to Use

Stop When Resolved

If the Problem Persists

If the APNEA PRESSURE alarm continues after fresh-gas supply, patient connection, breathing circuit leaks, hose or tube blockage, microbial filter condition, accessories, and ventilation settings have been ruled out, the issue may involve an internal ventilation, pressure sensing, gas delivery, or breathing system problem.

Remove the anesthesia machine from service. Label it Out of Service. Send it for repair or bench evaluation by qualified anesthesia service personnel.

Do not keep troubleshooting on an active patient.

Clinical Use Tip

Never troubleshoot an unresolved APNEA PRESSURE alarm while the patient depends on the anesthesia machine for ventilation. Move the patient to a backup ventilation method first, then investigate the machine, breathing circuit, and accessories.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Drager Apollo anesthesia machine reported APNEA PRESSURE alarm during use. Staff reported ventilation stopped or airway pressure was not being detected."

Cause

What was observed during troubleshooting.

Example:
"Found breathing circuit disconnected at patient Y-piece, preventing pressure buildup during ventilation."

Resolution

What action was taken.

Example:
"Reconnected breathing circuit, verified patient connection, checked fresh-gas flow, inspected hoses and microbial filter, ran checkout, and confirmed ventilation pressure returned with no repeat APNEA PRESSURE alarm. Device returned to service."

Helpful Details to Include (If Known)

Final Thought

The APNEA PRESSURE alarm must be treated as a ventilation safety issue first and a troubleshooting issue second. Confirm patient ventilation, then work logically through fresh gas, patient connection, leaks, blockages, accessories, and settings before suspecting internal failure. Clear documentation protects the patient, the clinical team, and the device history.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide