Drager Apollo

Alarm: CHECK AUX CGO

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Apollo

What This Guide Helps With

This guide helps troubleshoot the CHECK AUX CGO alarm on a Drager Apollo anesthesia machine when fresh-gas delivery or switching between the auxiliary common gas outlet and other ventilation modes may be unreliable.

The alarm may occur when switching over to Aux CGO or when switching from Aux CGO back to another ventilation mode. Because this can affect where fresh gas is delivered, the condition should be treated as a gas delivery and mode switching reliability issue.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Fresh-Gas Flow at the Aux CGO

Switch Aux CGO On and Off Several Times

If the Bag Does Not Inflate, Switch to the Internal Breathing System

If Switching Back From Aux CGO Fails, Repeat Aux CGO Switching

Use a Functional Outlet or Alternate Gas Delivery Path

Check External and Simple Causes

Determine Whether the Issue Is Accessory, Setup, or Device Related

Stop When Resolved

If the Problem Persists

If CHECK AUX CGO continues after fresh-gas flow, setup, tubing, outlet use, and Aux CGO switching have been checked, external and common causes have been ruled out.

A persistent CHECK AUX CGO alarm strongly indicates an internal Aux CGO switching, sensing, control, or gas-path fault.

The anesthesia machine should be removed from service, labeled Out of Service, and sent for repair or bench evaluation. Escalate to DragerService or qualified service support as appropriate.

Do not rely on Aux CGO if switching or fresh-gas delivery is inconsistent.

Clinical Use Tip

Never troubleshoot on an active patient. Move the patient to a backup device or approved alternate ventilation and gas delivery method first.

If fresh-gas delivery or mode switching is questionable, transfer care to a known functional system before continuing troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Drager Apollo anesthesia machine displayed CHECK AUX CGO alarm. Staff reported fault when switching to auxiliary common gas outlet and/or switching back from Aux CGO to another ventilation mode."

Cause

What was observed during troubleshooting.

Example:
"Verified CHECK AUX CGO alarm. Fresh-gas flow and Aux CGO switching were checked. External setup, tubing, bag connection, and outlet issues were inspected. Persistent alarm indicates likely internal Aux CGO switching or sensing fault."

Resolution

What action was taken.

Example:
"Machine removed from service and labeled Out of Service. Aux CGO not considered reliable for clinical use. Referred for repair or DragerService evaluation."

Helpful Details to Include (If Known)

Final Thought

The CHECK AUX CGO alarm should be treated as a gas delivery and switching reliability issue. Clinical Engineering should first verify the alarm, confirm fresh-gas flow, check external setup, and cycle Aux CGO as directed. If the alarm persists or switching remains unreliable, the correct action is to remove the anesthesia machine from service and escalate for repair.

Accurate CCR documentation protects patients, staff, and future troubleshooting.

That is successful troubleshooting.