Drager Apollo

Alarm: CIRCUIT LEAK

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Apollo

What This Guide Helps With

This guide helps troubleshoot a CIRCUIT LEAK alarm on a Drager Apollo anesthesia machine when a leak is detected in the patient circle system. A circuit leak can reduce delivered ventilation, affect tidal volume, interfere with pressure delivery, and compromise anesthetic gas delivery.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Expected outcome: The patient is protected from inadequate ventilation or gas delivery.

Why it matters: A patient circle system leak can prevent the anesthesia machine from delivering or maintaining the intended volume, pressure, or gas mixture.

2. Verify the Reported Alarm and Behavior

Expected outcome: The CIRCUIT LEAK alarm is confirmed and the leak behavior is clearly observed.

Why it matters: This confirms a true patient circle leak condition instead of a one-time setup issue or incorrect alarm report.

3. Remove From Patient Use if the Leak Cannot Be Immediately Corrected

Expected outcome: Clinical care continues without relying on a leaking breathing system.

Why it matters: A persistent circuit leak can create unsafe ventilation and anesthetic delivery conditions.

4. Check the Patient Breathing Circuit

Expected outcome: The breathing circuit is fully connected and free of obvious leaks or damage.

Why it matters: The patient circuit is the most common and easiest leak source to verify before suspecting the machine.

5. Check the Tube or Airway Connection

Expected outcome: The patient-side connection is secure and not allowing gas to escape.

Why it matters: A leak near the airway or test lung may appear as a patient circle system leak even when the machine itself is functioning correctly.

6. Check the Breathing Filter or HME

Expected outcome: The filter is correctly installed and not leaking.

Why it matters: Filters are frequently changed between patients and can cause leaks if not fully seated or if the housing is cracked.

7. Check All Adapters and Accessory Connections

Expected outcome: The circuit is simplified and all remaining connections are secure.

Why it matters: Every added adapter creates another possible leak point.

8. Check the Breathing Bag and Manual Ventilation Path

Expected outcome: The manual ventilation path holds pressure without obvious leakage.

Why it matters: A leak in the manual breathing path may be interpreted as a circle system leak depending on mode and setup.

9. Check the CO2 Absorber Canister and Breathing System Seating

Expected outcome: The absorber and breathing system seal properly.

Why it matters: A poorly seated absorber can create a significant circle system leak.

10. Perform a Leak Check or Checkout Test

Expected outcome: The Apollo passes leak testing and ventilates normally without the CIRCUIT LEAK alarm.

Why it matters: Passing a checkout confirms the leak was corrected before the anesthesia machine is returned to clinical use.

11. Isolate Accessory-Related vs Device-Related Leak

Expected outcome: The leak is isolated to a replaceable external component or remains with the anesthesia machine.

Why it matters: This prevents unnecessary repair escalation for a simple disposable circuit, filter, or adapter issue.

12. Stop When Resolved

If the Problem Persists

If the CIRCUIT LEAK alarm continues after the tube, hoses, filter, breathing bag, absorber canister, adapters, and external circuit components have been checked or replaced, external and common causes have been ruled out.

A persistent CIRCUIT LEAK alarm may indicate an internal breathing system seal issue, valve issue, internal pneumatic leak, sensor-related issue, or another device-level fault requiring bench evaluation.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot on an active patient longer than necessary. If ventilation is affected or the leak source is not immediately obvious, move the patient to a backup device or alternate approved ventilation method first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Drager Apollo anesthesia machine reported CIRCUIT LEAK alarm. Staff reported leak in patient circle system during setup/use with concern for breathing circuit leak."

Cause

What was observed during troubleshooting.

Example:
"Leak traced to loose/damaged external circuit component, breathing hose, tube connection, filter, breathing bag, absorber seal, or adapter connection."

Resolution

What action was taken.

Example:
"Checked patient breathing circuit, tube connection, hoses, filter, adapters, breathing bag, and absorber canister. Corrected/replaced leaking component and performed leak check. Alarm cleared and unit passed checkout. Returned to service."

Helpful Details to Include (If Known)

Final Thought

CIRCUIT LEAK alarms should be approached as patient safety issues first and equipment problems second. Start with the external breathing circuit, tube, hoses, filter, adapters, bag, and absorber before suspecting internal failure. If the leak cannot be corrected and verified by checkout, remove the anesthesia machine from service and document the findings clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide