Drager Apollo

Alarm: CO2 SENSOR FAIL CO2

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Apollo

What This Guide Helps With

This guide helps troubleshoot the CO2 SENSOR FAIL CO2 alarm on a Drager Apollo anesthesia machine when the onboard CO2 gas measurement system has failed or CO2 monitoring is unavailable. The alarm is listed for Standby only, but the condition should still be treated seriously because reliable CO2 monitoring is critical for anesthesia care, ventilation confirmation, and patient safety.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Confirm the Machine Is Not Being Relied On for CO2 Monitoring

Check for Other Related Gas Monitoring Alarms

Check External and Simple Causes First

Power Cycle Only If Safe and Appropriate

Verify External CO2 Monitoring If the Unit Must Remain Clinically Available

Determine Whether the Issue Is Accessory, Configuration, or Device Related

Stop When Resolved

If the Problem Persists

If the CO2 SENSOR FAIL CO2 alarm remains active or returns after external checks, accessory checks, and a safe restart, external and common causes have been ruled out.

A persistent alarm strongly indicates an internal CO2 gas measurement system failure or broader gas measurement system issue.

The anesthesia machine should be removed from service, labeled Out of Service, and sent for repair or bench evaluation. Escalate to DragerService or qualified anesthesia service support.

Do not continue using the device as a primary anesthesia machine without reliable CO2 monitoring or an approved external gas measuring system.

Clinical Use Tip

Never troubleshoot on an active patient. Move the patient to a backup device first. A device with an unresolved CO2 monitoring failure should not be trusted for patient ventilation monitoring until the issue is corrected or properly bypassed with approved external monitoring.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Drager Apollo anesthesia machine displayed CO2 SENSOR FAIL CO2 alarm in Standby. Staff reported onboard CO2 gas measurement was unavailable."

Cause

What was observed during troubleshooting.

Example:
"Confirmed CO2 SENSOR FAIL CO2 alarm. External sampling setup, visible connections, sample line, and water trap were checked with no correctable external issue found. Alarm persisted after safe restart, indicating likely internal CO2 gas measurement system failure."

Resolution

What action was taken.

Example:
"Removed anesthesia machine from service and labeled Out of Service. Advised use of approved external gas measuring system if clinically required. Referred unit for DragerService or qualified bench evaluation."

Helpful Details to Include (If Known)

Final Thought

CO2 SENSOR FAIL CO2 should be treated as a patient safety issue because reliable CO2 monitoring is essential for anesthesia care. Clinical Engineering should confirm the alarm, rule out simple external causes, protect the patient from unreliable monitoring, and escalate appropriately when the internal gas measurement system appears to have failed. Clear CCR documentation helps show what was checked, what was found, and why the device was removed from service.

That is successful troubleshooting.