Drager Apollo

Alarm: CONTINUOUS PRESSURE

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Apollo

What This Guide Helps With

This guide helps troubleshoot the CONTINUOUS PRESSURE alarm on a Drager Apollo anesthesia machine. This alarm indicates that breathing pressure has exceeded the set limit for more than 15 seconds. Possible causes include sustained high airway pressure, circuit obstruction, restricted exhalation, breathing system restriction, scavenging restriction, or an incorrect pressure alarm limit.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Remove From Patient Use if Pressure Does Not Clear

Check Ventilation and Patient Breathing Conditions

Check the Breathing Circuit and Hoses

Check the Breathing System Assembly

Check the Gas Scavenging System

Check Alarm Limit Settings

Test With a Known-Good Breathing Circuit or Test Lung

Stop When Resolved

If the Problem Persists

If the CONTINUOUS PRESSURE alarm continues after checking patient condition, ventilation settings, breathing hoses, breathing system setup, gas scavenging system, external accessories, and alarm limits, remove the anesthesia machine from service.

Label the unit Out of Service.

Send the anesthesia machine for repair, bench evaluation, or vendor service.

Persistent continuous pressure alarms may indicate an internal breathing system, valve, sensor, ventilator control, pressure monitoring, or pressure-relief-related failure.

Clinical Use Tip

Never troubleshoot a sustained pressure alarm for an extended time while the patient remains dependent on the anesthesia machine. Move the patient to a backup device or alternate ventilation method first if safe ventilation cannot be confirmed quickly.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Drager Apollo anesthesia machine reported CONTINUOUS PRESSURE alarm. Staff stated breathing pressure exceeded the alarm limit during ventilation and the alarm did not clear normally."

Cause

What was observed during troubleshooting.

Example:
"Breathing circuit inspection found a kinked expiratory hose causing restricted exhalation and sustained breathing pressure above the set alarm limit."

Resolution

What action was taken.

Example:
"Removed kinked circuit tubing, installed proper breathing circuit setup, verified ventilation with test lung, confirmed pressure returned to expected range, and confirmed CONTINUOUS PRESSURE alarm did not return. Unit returned to service."

Helpful Details to Include (If Known)

Final Thought

A CONTINUOUS PRESSURE alarm should be treated as a real ventilation safety concern until proven otherwise. Start with the patient, then check the breathing circuit, breathing system, scavenging path, and alarm limits before assuming internal machine failure. If the alarm persists after external causes are ruled out, remove the device from service and escalate appropriately. Clear documentation protects the patient, the clinical team, and the technician.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide