What This Guide Helps With

This guide helps troubleshoot the FAN FAIL alarm on a Drager Apollo anesthesia machine. This alarm indicates the fan for evacuating gases inside the device may be defective. If fan failure occurs with internal leakage, oxygen concentration inside the anesthesia machine may rise and create a fire risk.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Do Not Attempt Continued Operation

Power Down the Machine Safely

Check for Obvious External Conditions Only

Verify Power Source and Environment

Do Not Open the Machine for Internal Fan Troubleshooting

Remove From Service

If the Problem Persists

If the FAN FAIL alarm remains active, repeats after restart, or cannot be clearly explained by an obvious external obstruction, external and simple causes have been ruled out.

A persistent FAN FAIL alarm may indicate an internal fan, fan control, wiring, sensing, or internal ventilation system problem.

The device should be removed from service, labeled Out of Service, and sent for repair or bench evaluation. Escalate to DragerService or qualified anesthesia service support. This is not an alarm to work around.

Clinical Use Tip

Never troubleshoot a FAN FAIL alarm on an active patient unless clinical staff determine it is necessary for immediate patient safety. Move the patient to a backup anesthesia machine or safe ventilation method before extended troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Drager Apollo anesthesia machine displayed FAN FAIL alarm. Staff reported concern that the fan for evacuating gases inside the device may be defective."

Cause

What was observed during troubleshooting.

Example:
"Confirmed FAN FAIL alarm. External vents, power source, and visible machine condition were checked. No safe external correction was available. Alarm indicates possible internal fan or internal ventilation system fault requiring service evaluation."

Resolution

What action was taken.

Example:
"Removed anesthesia machine from patient use, powered unit down after safe clinical transition, labeled Out of Service, and escalated for DragerService or qualified repair evaluation."

Helpful Details to Include (If Known)

Final Thought

FAN FAIL on a Drager Apollo should be treated as a serious safety alarm because internal gas evacuation may be impaired and fire risk may increase if internal leakage is present. The safest approach is to protect the patient first, confirm the alarm, perform only safe external checks, power the unit down, remove it from service, and escalate appropriately.

That is successful troubleshooting.