Drager Apollo

Alarm: O2 FLOW MEAS. FAIL

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Apollo

What This Guide Helps With

This guide helps troubleshoot the O2 FLOW MEAS. FAIL alarm on a Drager Apollo anesthesia machine when the oxygen fresh-gas flow measurement has failed or cannot be trusted.

This alarm affects the machine’s ability to reliably measure O2 fresh-gas flow. Even if gas may still be flowing, the displayed O2 flow measurement may not be dependable, which can affect clinical confidence during setup, checkout, or active use.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Remove the Machine From Patient Use if the Alarm Is Active and Unresolved

Use Oxygen Only as Fresh Gas for Initial Isolation

Compare Fresh-Gas Flow Settings to the Total Flow Meter

Confirm Fresh-Gas Flow Is Adequate for Ventilation Needs

Check External Gas Supply Conditions

Check for Setup or Configuration Issues

Power Cycle Only if Safe and Appropriate

Determine Whether the Issue Is External, Configuration-Related, or Device-Related

Stop When Resolved

If the Problem Persists

If O2 FLOW MEAS. FAIL remains active or returns after external oxygen supply, fresh-gas flow settings, total flow verification, setup checks, and safe restart have been completed, the likely issue is internal O2 fresh-gas flow measurement failure.

At that point, external causes have been ruled out. Remove the Apollo from service, label it Out of Service, and do not use it for patient care.

Send the machine for repair or bench evaluation. Escalate to Drager service or qualified anesthesia service support. Persistent O2 fresh-gas measurement failure is not a bedside troubleshooting issue.

Clinical Use Tip

Never troubleshoot an unresolved oxygen flow measurement alarm on an active patient. Move the patient to a backup anesthesia machine or safe ventilation method first if oxygen flow cannot be confidently verified.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Drager Apollo anesthesia machine reported O2 FLOW MEAS. FAIL during setup. Staff reported the O2 fresh-gas flow measurement was not reliable."

Cause

What was observed during troubleshooting.

Example:
"Checked oxygen pipeline connection, oxygen supply hose, flow settings, and total flow meter response. Alarm persisted after external checks, indicating likely internal O2 fresh-gas flow measurement fault."

Resolution

What action was taken.

Example:
"Removed Apollo anesthesia machine from service and labeled Out of Service. Advised staff not to use for patient care. Unit requires bench evaluation and Drager service support for persistent O2 FLOW MEAS. FAIL alarm."

Helpful Details to Include (If Known)

Final Thought

O2 FLOW MEAS. FAIL should be treated as a real oxygen flow verification problem until proven otherwise. Start with patient safety, verify the alarm, check oxygen supply and flow settings, use the total flow meter for practical confirmation, and escalate when the alarm persists.

Good troubleshooting protects the patient, supports the clinician, and gives Clinical Engineering a clear stopping point when the issue points to an internal measurement failure.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide