Drager Apollo

Alarm: PEEP HIGH

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Apollo

What This Guide Helps With

This guide helps troubleshoot a PEEP HIGH alarm on a Drager Apollo anesthesia machine. The alarm indicates that expiratory pressure is elevated above the expected PEEP level, which may be caused by ventilation settings, patient or airway resistance, a restricted breathing circuit, filter blockage, scavenging restriction, or device-related pressure control issues.

The reported condition is that expiratory pressure is 5 cmH2O above PEEP for 2 breaths, or 5 cmH2O above PEEP in Pressure Support mode for more than 30 seconds.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Ventilation Parameters

Check for Patient or Airway Causes

Inspect the Breathing Circuit

Check the Expiratory Valve Area and Manual/Automatic Ventilation Setup

Check the Anesthetic Gas Scavenging Line

Check External Accessories and Filters

Run a Pre-Use or System Check

Isolate Whether the Issue Is External or Device-Related

Stop When Resolved

If the Problem Persists

If the PEEP HIGH alarm continues after ventilation settings, patient causes, breathing circuit, filters, scavenging line, and external accessories have been ruled out, remove the anesthesia machine from service.

Persistent PEEP HIGH alarms may indicate an internal expiratory valve, pressure sensing, ventilation control, breathing system, or scavenging interface problem.

Clinical Use Tip

Never troubleshoot repeated PEEP HIGH alarms on an active patient unless anesthesia staff confirm it is safe. Elevated expiratory pressure can affect ventilation and patient safety. Move the patient to a backup device or alternate ventilation support before extended troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Drager Apollo anesthesia machine displayed PEEP HIGH alarm during automatic ventilation. Expiratory pressure remained above the set PEEP level during use."

Cause

What was observed during troubleshooting.

Example:
"Ventilation parameters and breathing circuit were checked. Scavenging line was inspected and found restricted, creating excess pressure during exhalation."

Resolution

What action was taken.

Example:
"Corrected scavenging line routing and verified proper connection. Replaced questionable external tubing as needed. Ran machine check and confirmed PEEP HIGH alarm did not return. Device returned to service."

Helpful Details to Include (If Known)

Final Thought

A PEEP HIGH alarm should be treated as a real ventilation concern until proven otherwise. Start with patient safety, confirm the alarm behavior, check ventilation settings, inspect the breathing circuit, and verify the scavenging line before suspecting internal failure. Clear documentation helps show whether the cause was external, clinical, or device-related.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide