Drager Apollo

Alarm: PINSP NOT ACHIEVED

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Apollo

What This Guide Helps With

This guide helps troubleshoot the PINSP NOT ACHIEVED alarm on a Drager Apollo anesthesia machine when the set inspiratory pressure in Pressure Mode is not being reached. Possible causes include ventilation settings, breathing circuit leaks, patient connection leaks, accessory problems, or fresh-gas shortage.

This alarm can affect delivered ventilation and should be handled with patient safety first, especially during active anesthesia use.

Step-by-Step Troubleshooting

Ensure Patient Safety First

The goal is to keep the patient safely ventilated while the alarm condition is evaluated. PINSP NOT ACHIEVED can indicate that the machine is not reaching the pressure needed for the selected ventilation mode.

Verify the Reported Alarm and Behavior

This helps confirm a true ventilation pressure issue instead of a misunderstood setting, temporary disconnect, or setup problem.

Check the Set Ventilation Parameters

If the set pressure, timing, or mode is inappropriate, the machine may not achieve the expected inspiratory pressure even though the ventilator is functioning normally. If correcting the ventilation settings clears the alarm and ventilation is stable, stop troubleshooting and document the corrected setting issue.

Check for Breathing Circuit Leaks

A leak is one of the most likely causes of PINSP NOT ACHIEVED because pressure cannot build if gas escapes from the breathing system. If a loose or damaged circuit component is corrected and the alarm clears, verify ventilation stability before returning the machine to use.

Check the Patient Connection and Airway Interface

The machine may be working properly, but a leak at the patient airway can prevent the set inspiratory pressure from being achieved.

Check Fresh-Gas Flow and Gas Supply

A fresh-gas shortage can prevent the system from maintaining the volume and pressure needed to achieve the set inspiratory pressure. If increasing or correcting fresh-gas flow clears the alarm, confirm stable ventilation and document the setting-related cause.

Perform a Leak Test or System Check

A system leak test helps separate external circuit or setup problems from a machine-side leak or internal ventilator issue.

Isolate Accessory-Related Causes

Many pressure alarms are caused by disposable accessories or setup issues rather than internal machine failure.

Check for Physical Damage, Contamination, or Misassembly

Damaged seals or improperly seated parts can create leaks large enough to prevent inspiratory pressure from being achieved.

Determine Whether the Issue Is External or Device-Related

The issue should be narrowed to settings, accessory, gas supply, leak, or machine failure before deciding the next action.

Stop When Resolved

If the Problem Persists

If PINSP NOT ACHIEVED continues after ventilation settings, patient circuit, accessories, fresh-gas flow, external leaks, and system checkout have been ruled out, the issue may indicate an internal ventilator, breathing system, flow delivery, pressure sensing, or control problem.

Remove the anesthesia machine from service. Label it Out of Service. Send it for repair, service evaluation, or qualified bench testing.

Do not leave the machine available for patient use if it cannot reliably achieve set inspiratory pressure.

Clinical Use Tip

Never troubleshoot on an active patient. Move patient to backup device first.

Do not troubleshoot an unresolved ventilation pressure alarm on an active patient unless clinical staff are maintaining safe ventilation. If inspiratory pressure is not being achieved, patient ventilation must be stabilized first using manual ventilation or another anesthesia machine if needed.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Drager Apollo displayed PINSP NOT ACHIEVED alarm during Pressure Mode ventilation. Set inspiratory pressure was not being reached."

Cause

What was observed during troubleshooting.

Example:
"Breathing circuit leak found at loose patient filter connection. Fresh-gas supply and ventilation settings were checked and no gas supply issue was found."

Resolution

What action was taken.

Example:
"Reseated filter and circuit connection. Verified alarm cleared, ventilation pressure stabilized, and machine passed checkout. Returned to service."

Helpful Details to Include (If Known)

Final Thought

PINSP NOT ACHIEVED should be treated as a ventilation performance alarm, not just a nuisance message. Start with patient safety, confirm the alarm, check settings, leaks, accessories, and fresh-gas supply first, then escalate when the machine still cannot achieve pressure. Clear documentation helps show what was ruled out and why the device was returned to service or removed for repair.

That is successful troubleshooting.