Drager Apollo

Alarm: POWER FAIL

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Apollo

What This Guide Helps With

This guide helps troubleshoot the POWER FAIL alarm on a Drager Apollo anesthesia machine. This alarm may occur when AC power is lost, the machine is operating on battery, or a device connected to an auxiliary outlet has caused a short-circuit condition.

A POWER FAIL alarm can affect ventilation support, gas delivery readiness, monitoring, and powered accessories. Treat the alarm as a patient safety issue first, then work through external power and accessory checks before suspecting an internal machine fault.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Main AC Power First

Observe Battery Status

Check for Auxiliary Outlet Overload or Short-Circuit

Verify Pipeline Supply and Machine Readiness

Check External Power Cord and Physical Condition

Retest After External Causes Are Removed

If the Problem Persists

If the POWER FAIL alarm continues after AC power has been verified, the outlet has been checked, the power cord has been inspected, auxiliary outlet devices have been unplugged, battery status has been reviewed, and external damage has been ruled out, the issue may involve an internal power supply, auxiliary outlet circuit, battery charging circuit, power distribution issue, or internal machine fault.

Clinical Use Tip

Never troubleshoot a POWER FAIL alarm on an active patient longer than necessary. Move the patient to a safe backup device or backup ventilation plan first if power cannot be restored quickly. Do not rely on an anesthesia machine with unresolved power failure alarms.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Drager Apollo anesthesia machine displayed POWER FAIL alarm. Staff reported the machine was operating on battery and the alarm occurred after an accessory was connected to an auxiliary outlet."

Cause

What was observed during troubleshooting.

Example:
"POWER FAIL alarm indicates AC power loss or possible short-circuit from a device connected to an auxiliary outlet. External power source, power cord, battery status, and auxiliary outlet devices were checked. Fault was isolated to a connected auxiliary device."

Resolution

What action was taken.

Example:
"Verified patient safety, checked AC outlet, power cord, battery status, pipeline supply, and auxiliary outlet accessories. Disconnected suspected auxiliary device and retested. POWER FAIL alarm cleared and normal AC power operation was confirmed before return to service."

Helpful Details to Include (If Known)

Final Thought

A POWER FAIL alarm on a Drager Apollo should be treated as a patient safety issue first and a power isolation problem second. Start with AC power, battery condition, connected accessories, and auxiliary outlet devices before suspecting internal failure. If the alarm persists after external causes are ruled out, remove the machine from service and escalate for repair. Clear documentation helps show what was checked, what was ruled out, and why the final decision was made.

That is successful troubleshooting.