Drager Apollo

Alarm: POWER SPLY ERROR

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Apollo

What This Guide Helps With

This guide helps troubleshoot the POWER SPLY ERROR alarm on a Drager Apollo anesthesia machine. This alarm indicates an internal fault in the power supply. The machine may continue operating for the time being, but the alarm should still be treated as a reliability concern because power system faults can affect continued safe operation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check the External Power Source

Inspect the Power Cord and Plug

Check Machine Position and Cable Strain

Check for Obvious Environmental or Physical Causes

Confirm Battery and Charging Behavior

Power Cycle Only When Clinically Safe

Determine Whether the Issue Is External or Device-Related

If the Problem Persists

If POWER SPLY ERROR remains active or returns after the external power source, power cord, cable strain, battery behavior, and basic restart checks have been completed, remove the Apollo from service. The manual identifies this condition as an internal power supply fault, even though operation may continue for the time being.

Clinical Use Tip

Never troubleshoot on an active patient in a way that could interrupt ventilation, gas delivery, vapor delivery, or monitoring. If the machine needs to be power cycled, removed, or evaluated further, move the patient to a backup anesthesia machine or manual ventilation setup first under anesthesia direction.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Drager Apollo anesthesia machine displayed POWER SPLY ERROR. Staff reported the machine continued operating, but the alarm indicated a power supply fault."

Cause

What was observed during troubleshooting.

Example:
"Verified POWER SPLY ERROR condition. External AC outlet, power cord seating, cord condition, cable strain, battery status, and obvious external damage were checked. Alarm remained consistent with internal power supply fault."

Resolution

What action was taken.

Example:
"Removed anesthesia machine from service, labeled Out of Service, and referred for repair/bench evaluation. DragerService support required for internal power supply fault."

Helpful Details to Include (If Known)

Final Thought

POWER SPLY ERROR should be handled as a reliability warning, not ignored because the Apollo may continue operating temporarily. Patient safety comes first, external power problems should be ruled out, and persistent power supply alarms should be escalated for service with clear CCR documentation.

That is successful troubleshooting.