Drager Apollo

Alarm: PRESSURE LIMITING

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Apollo

What This Guide Helps With

This guide helps troubleshoot a PRESSURE LIMITING alarm on a Drager Apollo anesthesia machine. This alarm indicates the ventilator is operating with pressure limitation. Possible causes include ventilation settings, a kinked or stenosed tube, restricted breathing hoses, or a soiled microbial filter on the inspiration side.

The main concern is that pressure limitation may prevent the ventilator from delivering the expected breath. Treat this as a ventilation safety issue first, then work through external causes before assuming internal machine failure.

Step-by-Step Troubleshooting

Ensure Patient Safety First

The expected outcome is that patient ventilation is maintained without relying on a machine with an unresolved pressure limitation condition.

Verify the Reported Alarm and Behavior

The expected outcome is that the alarm and trigger conditions are clearly understood before parts or settings are changed.

Check the Ventilation Settings

If the alarm resolves after anesthesia corrects an improper setting, document the setting-related cause and verify normal operation before return to use.

Inspect the Breathing Circuit and Patient Tubing

The expected outcome is that the tubing is open, correctly connected, and free of restriction.

Check the Inspiratory-Side Microbial Filter

A soiled or saturated microbial filter can increase resistance and cause the ventilator to operate under pressure limitation.

Check for Obstruction at the Patient Connection

The expected outcome is that there is no downstream restriction between the machine and the patient airway.

Check the Breathing System Setup

Incorrect circuit assembly or accessory placement can create resistance that appears as a ventilator pressure limitation problem.

Isolate Accessory-Related Causes

The expected outcome is that the alarm clears after the restricted or incorrect accessory is replaced.

Perform a Basic Machine Check After Corrections

Clearing the alarm is not enough. The machine should be verified under a controlled checkout before being returned to patient care.

If the Problem Persists

If the PRESSURE LIMITING alarm continues after ventilation settings, tubing, hoses, patient connection, breathing circuit setup, and microbial filters have been checked, external and common causes have been ruled out.

A persistent pressure limitation condition may indicate an internal ventilator control, pressure sensing, flow delivery, valve, or breathing system issue.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot on an active patient when the troubleshooting could interrupt ventilation, pressure delivery, or patient monitoring. Move the patient to a backup anesthesia machine or alternate ventilation method first when the alarm cannot be corrected immediately and safely.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Drager Apollo reported PRESSURE LIMITING alarm during ventilation. Staff reported ventilator was operating with pressure limitation and requested Clinical Engineering evaluation."

Cause

What was observed during troubleshooting.

Example:
"Inspiratory microbial filter was found soiled and restricted, causing increased inspiratory resistance and repeated PRESSURE LIMITING alarm."

Resolution

What action was taken.

Example:
"Removed device from patient use for inspection. Checked ventilation settings, breathing hoses, patient tubing, breathing circuit setup, and inspiratory microbial filter. Replaced restricted microbial filter. Verified alarm cleared and machine passed required checkout before return to service."

Helpful Details to Include (If Known)

Final Thought

The PRESSURE LIMITING alarm should be approached as a ventilation safety issue first. Start with the patient, then verify the alarm, then check settings, tubing, hoses, filters, and circuit setup before assuming internal failure. If the alarm persists after external causes are ruled out, remove the machine from service and escalate for proper evaluation. Clear documentation protects the patient, the clinical team, and the equipment history.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide