Drager Apollo Anesthesia Machine

Fails Automatic Self-Test

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Apollo

What This Guide Helps With

This guide addresses situations where the Apollo anesthesia machine:

The focus is on logical, external troubleshooting steps first before assuming internal failure.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Self-test failures indicate safety-critical issues. Do not attempt to bypass startup checks.

2. Confirm AC Power Source

Why this matters:
Low or unstable input power can cause incomplete self-tests or false failure messages.

3. Verify Gas Supply Pressures

Why this matters:
The Apollo performs gas system integrity checks during startup. Low or absent supply pressure will cause test failure.

4. Check Vaporizers

Why this matters:
Improperly seated vaporizers can trigger leak or configuration errors.

5. Inspect Breathing Circuit Assembly

Why this matters:
Leak tests are part of the automatic self-test. An improperly assembled circuit will cause failure.

6. Check Water Trap and Flow Sensors

Why this matters:
Moisture accumulation can interfere with flow measurement and sensor calibration checks.

7. Restart the System

8. Remove Optional Accessories

Why this matters:
Peripheral communication errors can cause startup faults.

If the Problem Persists

If:

The issue is likely internal (sensor calibration failure, internal leak, power supply module fault, or board-level issue).

At this point:

Knowing when to stop external troubleshooting prevents unnecessary risk and protects patient safety.

Clinical Use Tip

Never troubleshoot a failed self-test on an active patient.

If the failure occurs during room setup:

Self-tests are safety-critical on anesthesia delivery systems.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Anesthesia machine failed automatic self-test during morning setup. Will not enter ready mode.”

Cause

What was observed during troubleshooting.

Example:
“Self-test failure traced to improperly seated vaporizer causing leak test error.”

Resolution

What action was taken.

Example:
“Reseated and locked vaporizer properly. Restarted system. Self-test completed successfully. Device returned to service.”

Helpful Details to Include

Final Thought

An anesthesia machine self-test failure should always be approached methodically. Start with power, gas supply, vaporizers, and circuit assembly before assuming internal malfunction. Protect patient safety by removing the device from service when necessary, and document findings clearly to support continuity of care and technical accountability.

That is successful troubleshooting.