Drager Apollo Anesthesia Machine

Flow Sensor Calibration Errors

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Apollo

What This Guide Helps With

This guide addresses situations where the Apollo anesthesia machine reports flow sensor calibration errors, such as alarms during startup or spontaneous calibration failure during operation. These errors may impact accurate delivery of tidal volume and anesthetic gases.

The guide focuses on external checks and setup verification first, before assuming internal sensor or module failure.

Step-by-Step Troubleshooting

Confirm Machine Power and Startup

Check Gas Supply and Pressures

Inspect Flow Sensor Pathway

Check Breathing Circuit Configuration

Run Automatic Calibration Again

Swap Sensors if Possible

Inspect Environmental Conditions

If the Problem Persists

External checks have been completed and calibration still fails.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Flow sensor calibration fails during startup; machine shows error FS-03.”

Cause

What was observed during troubleshooting.

Example:
“Observed condensation in the flow sensor and loose sensor connector. Calibration failed on both adult and pediatric sensors.”

Resolution

What action was taken.

Example:
“Cleared condensation, reseated sensor connectors, re-ran calibration successfully. Machine passed self-test and ready for use.”

Helpful Details to Include

Final Thought

Flow sensor calibration errors often stem from external factors such as moisture, gas supply issues, or misalignment. Systematic verification avoids unnecessary downtime and maintains patient safety. Proper documentation and escalation are critical when external checks do not resolve the issue.

That is successful troubleshooting.