On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Flow-sensor calibration failures caused by contamination, moisture, incorrect installation, damaged removable sensors, circuit conditions, or external connection problems.
Step-by-Step Troubleshooting
1. Protect the Patient Before Troubleshooting
Do not troubleshoot a flow-measurement or calibration problem while the anesthesia machine is being relied upon for patient ventilation or monitoring. Move the patient to another verified system first.
Expected outcome: Patient ventilation and monitoring are maintained independently.
2. Confirm the Calibration Failure
Repeat the approved calibration or system test under controlled conditions and record the exact displayed message and affected sensor or channel.
Determine whether the problem is consistent or intermittent.
Expected outcome: The calibration failure is reproducible and clearly identified.
3. Inspect the Flow Sensor Installation
Verify the externally accessible flow sensor is the correct type, fully seated, correctly oriented where applicable, and installed without mechanical interference.
Do not disassemble the sensor beyond normal removable components.
Expected outcome: The sensor is correctly installed. If reseating it allows successful calibration, troubleshooting can stop after verification.
4. Inspect for Moisture or Contamination
Examine accessible sensor surfaces and associated breathing-system components for condensation, secretions, absorbent dust, debris, or other contamination.
Follow approved handling and cleaning practices for the removable component.
Expected outcome: The sensing pathway is clean and dry enough for normal operation.
5. Check Associated Breathing-System Components
Inspect nearby breathing-system connections, tubing, valves, circuit components, and seals for incorrect assembly, leaks, or obstruction that could interfere with calibration conditions.
Expected outcome: The surrounding breathing system is correctly assembled and free of obvious external defects.
6. Substitute a Known-Good Compatible Sensor
If permitted and available, install a compatible known-good flow sensor and repeat the calibration or system test.
Use the substitution to distinguish an accessory problem from a machine-side problem.
Expected outcome: Calibration succeeds with the known-good sensor, identifying the original removable sensor as the cause.
7. Verify the Machine Completes the Relevant System Test
After correcting external conditions, repeat the required system or sensor calibration test.
Do not return the machine to service if calibration continues to fail.
Expected outcome: Flow-sensor calibration completes normally without recurring errors.
8. Verify Ventilation and Volume Measurement
Connect an appropriate test lung and confirm stable ventilation, reasonable inspired and expired volume measurements, and normal related alarm behavior.
Use approved test equipment when quantitative verification is required.
Expected outcome: Flow-derived measurements and ventilation behavior are stable after calibration.
9. Escalate if Calibration Still Fails
If calibration continues to fail with proper installation, a clean and dry breathing system, and a known-good sensor, stop external troubleshooting.
Expected outcome: The machine is removed from clinical use for bench evaluation.
If the Problem Persists
Common external causes have been ruled out. The remaining condition may involve the sensor interface, internal measurement pathway, pneumatic system, signal processing, calibration circuitry, or service-level configuration.
Remove the machine from service, label it Out of Service, and evaluate it using appropriate Drager service documentation and approved test equipment. Internal repairs and service-level calibration should be performed only by qualified personnel.
Complete calibration, system testing, ventilation verification, and relevant alarms before returning the machine to service.
Knowing when to stop external troubleshooting and escalate is proper troubleshooting.
Clinical Use Tip
Do not rely on displayed volume information when flow-sensor calibration is unsuccessful; use another verified anesthesia machine for patient care.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"OR staff reported the Atlan A300 repeatedly failed flow-sensor calibration during the system test."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found moisture on the removable flow sensor and adjacent breathing-system component."
Resolution
What action was taken.
Example:
"Corrected the moisture condition using approved handling procedures, reinstalled the sensor, completed calibration and the system test successfully, and verified ventilation with a test lung."
Helpful Details to Include (If Known)
- Exact calibration message
- Sensor location or channel affected
- Sensor seating and orientation
- Moisture or contamination observed
- Breathing-system condition
- Known-good sensor substitution
- System-test result
- Volume measurements after correction
- Alarm behavior
- Final device status
Final Thought
Flow-sensor calibration failures should be approached through installation, cleanliness, breathing-system setup, and known-good substitution before internal measurement failures are considered.
That is successful troubleshooting.