On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Communication failures caused by disconnected cables, incorrect ports, network infrastructure, peripheral devices, configuration, or downstream integration problems.
Step-by-Step Troubleshooting
1. Protect the Patient and Preserve Required Monitoring
Determine whether the communication problem affects only documentation/data transfer or also removes clinically required remote monitoring, alarms, or integrated displays.
If required clinical information or alarm visibility is lost, provide an alternate verified monitoring method before troubleshooting.
Expected outcome: Patient monitoring and alarm coverage remain adequate despite the communication failure.
2. Define the Failed Communication Path
Identify exactly what is not working: network connectivity, data export, external monitor communication, central monitoring integration, charting interface, or another connected system.
Determine whether the failure affects one machine, one room, one destination, or multiple devices.
Expected outcome: The affected communication path and endpoints are clearly identified.
3. Inspect External Cables and Connections
Check Ethernet, serial, USB, display, or other externally accessible communication cables involved in the reported function. Verify connectors are fully seated and cables are free of cuts, crushed sections, bent contacts, or strain.
Expected outcome: Physical connections are secure and undamaged. If communication returns after correcting a loose cable, troubleshooting can stop after end-to-end verification.
4. Verify the Correct Port and Connected Device
Confirm the cable is connected to the intended machine port and correct downstream device or network jack. Check for recent room changes, cable swaps, or equipment replacements that could have changed the connection path.
Expected outcome: Each endpoint is connected to the intended physical interface.
5. Check Link and Status Indicators
Observe accessible network or communication status indicators on the machine, connected equipment, and wall/network interface when available.
Do not interpret the absence of a link indicator as proof of a machine hardware failure until the cable and infrastructure are tested.
Expected outcome: The physical communication link is active, or the failed segment is narrowed.
6. Substitute Known-Good External Components
When appropriate, test with a known-good compatible network cable, interface cable, port, wall jack, or connected peripheral.
Change one component at a time when practical.
Expected outcome: Communication returns after substitution, identifying the failed external cable, port, or peripheral.
7. Verify User-Accessible Communication Settings
Confirm normal user-accessible network or communication settings have not been unintentionally changed. Compare with a functioning peer device when appropriate.
Do not alter protected network parameters, integration settings, or service configuration without authorization and documented change control.
Expected outcome: The visible configuration is consistent with the intended installation.
8. Test the Entire Communication Path
Verify whether information leaves the Atlan A300 and reaches the intended receiving system. When relevant, coordinate with clinical informatics, networking, integration, or monitoring personnel to determine where the data path stops.
Expected outcome: The failing segment is isolated to the machine, cabling, network infrastructure, integration interface, or receiving system.
9. Verify Local Anesthesia Functions Remain Normal
Complete the normal machine system test and confirm the communication problem has not affected ventilation, gas delivery, controls, alarms, or local monitoring.
Expected outcome: Local anesthesia-machine functions remain normal and independent of the communication fault.
10. Perform End-to-End Verification After Correction
After repairing a connection or restoring configuration/infrastructure, verify that the expected patient or device data appears at the intended destination and that the connection remains stable.
If the communication path includes alarm or monitoring information, verify it at the receiving station.
Expected outcome: Data communication is restored through the complete intended path.
11. Escalate Unresolved Communication Failures
If known-good cables, ports, network connections, and authorized configuration checks do not resolve the issue, stop external troubleshooting and involve the appropriate support group.
Expected outcome: The device or infrastructure issue is appropriately escalated without unnecessary configuration changes.
If the Problem Persists
Common external causes have been ruled out. The remaining issue may involve an internal network interface, communication hardware, software, protected configuration, hospital network infrastructure, integration middleware, interface engine, or receiving-system configuration.
If the machine itself is suspected, remove it from service when the failed communication function is required for safe clinical use, label it Out of Service, and evaluate it using appropriate Drager documentation and approved test equipment. Infrastructure and integration issues should be escalated to the appropriate networking, clinical informatics, or monitoring support team.
Any repair or configuration change should be performed only by qualified and authorized personnel. Verify both local machine operation and the complete communication path before return to service.
Knowing when to stop external troubleshooting and escalate is proper troubleshooting.
Clinical Use Tip
When integrated monitoring fails, verify the information and alarms at the actual receiving station rather than assuming that a local network indicator proves end-to-end communication.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"OR staff reported the Atlan A300 was operating normally but anesthesia data was no longer reaching the integrated monitoring system."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found the network cable partially disconnected from the machine's external communication connection."
Resolution
What action was taken.
Example:
"Reseated the cable, confirmed the physical link, verified data at the receiving monitoring system, completed a local functional check, and returned the machine to service."
Helpful Details to Include (If Known)
- Specific communication function affected
- Receiving system or destination
- Room and network jack
- Cable and connector condition
- Link or status indicators
- Known-good cable or port tested
- Whether other devices were affected
- Visible configuration compared with peer device
- Receiving-system verification
- Local machine functional status
- Results before and after correction
- Final device status
Final Thought
Communication troubleshooting should follow the complete path from the anesthesia machine to the receiving system, verify physical causes before configuration changes, and preserve required monitoring throughout the process.
That is successful troubleshooting.