Drager Atlan A350

Network Time, Device Configuration, or Case Data Will Not Synchronize

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Atlan A350

What This Guide Helps With

Troubleshoot network connectivity, incorrect synchronized time, configuration distribution, case-data transfer, cabling, infrastructure, and external interface problems.

Step-by-Step Troubleshooting

1. Protect the Patient and Preserve Clinical Workflow

Do not interrupt an active anesthetic solely to troubleshoot network, time, configuration, or case-data synchronization unless the issue creates a separate patient-safety concern.

If the machine is otherwise clinically functional:

Expected outcome: Patient care and required clinical documentation continue safely while the synchronization issue is isolated.

2. Confirm What Is Actually Failing

Determine whether the reported problem involves:

Check whether the condition affects only one Atlan A350 or multiple devices.

Expected outcome: The failure is narrowed to time, configuration, network transport, case data, or a receiving-system problem.

If the reported function is currently working normally and repeated testing shows no fault, document the findings and stop troubleshooting.

3. Check Network Cable and Physical Connection

Inspect the accessible network connection.

Check for:

Reseat accessible network connections when safe.

Expected outcome: The physical network connection is secure and undamaged.

If reseating the cable restores synchronization and communication remains stable, proceed to final verification and stop troubleshooting.

4. Test With a Known-Good Network Cable

If the cable is suspect, substitute a known-good cable appropriate for the network environment.

Avoid changing multiple infrastructure components simultaneously.

Expected outcome: The device establishes normal network connectivity with the known-good cable.

If communication returns, replace the defective cable, document the cause, and complete final verification.

5. Verify the Network Port or Connection Path

Determine whether the connected wall jack, switch path, or approved network connection is operational.

Where facility policy permits:

Do not move the anesthesia machine to an arbitrary network segment as a workaround.

Expected outcome: The connected network port provides the expected authorized connectivity.

If the infrastructure port is the cause, coordinate correction with IT and verify synchronization afterward.

6. Review Device Network Status Without Unauthorized Changes

Use normal user-accessible status screens to review available connection information and indications.

Look for:

Record existing settings before any approved change.

Do not alter protected network parameters, addresses, certificates, security settings, or service configuration without authorization.

Expected outcome: The device shows a network state consistent with the known facility configuration.

If an approved correction restores communication, proceed to functional verification.

7. Determine Whether the Problem Is Device-Specific or Infrastructure-Wide

Compare the affected Atlan A350 with another similar functioning device or another endpoint on the same supported workflow when available.

Ask:

Expected outcome: The fault is isolated to the machine, local connection, network infrastructure, or receiving application.

If multiple devices are affected, escalate the infrastructure issue rather than performing unnecessary service on each anesthesia machine.

8. Verify Time Synchronization Behavior

If the reported issue is incorrect time:

Do not manually compensate for persistent synchronization failure unless permitted by approved policy and configuration procedures.

Expected outcome: The Atlan A350 displays the correct facility time and remains synchronized.

If correct time synchronization returns after restoring network connectivity, the issue is resolved.

9. Verify Case Data and Receiving-System Availability

If network connectivity appears normal but case data does not transfer, determine where the data flow stops.

Verify:

Protect patient information during testing and use approved test records where required.

Expected outcome: Case data traverses the expected workflow or the failure is isolated to a specific downstream system.

If the Atlan A350 sends data normally and the receiving system remains unable to process it, escalate to the appropriate IT, integration, or application support group.

10. Review Approved Configuration Consistency

If the problem involves configuration synchronization, compare approved user-visible configuration with a known-good equivalent device when appropriate.

Check for:

Do not copy protected settings or make unauthorized service-level configuration changes.

Expected outcome: Approved device configuration is consistent with the intended clinical environment.

If an authorized configuration correction restores synchronization, document the change and verify all affected functions.

11. Use a Controlled Restart Only When Clinically Safe

If the machine is off-patient and approved workflow allows it, perform a normal controlled shutdown and restart.

Do not remove power abruptly to resolve a communications problem.

After restart, observe whether:

Expected outcome: The system completes startup normally and all required network functions reconnect.

If synchronization returns and remains stable, proceed to final verification.

12. Perform Final End-to-End Verification

Before closing the work order, verify the affected function from source to destination.

As applicable:

Expected outcome: The complete communication path functions normally and consistently.

If the end-to-end path is verified, troubleshooting can stop.

If the Problem Persists

If power, physical cabling, network connection, local port, user-accessible configuration, device restart, receiving-system availability, and basic infrastructure checks have been completed, the remaining problem may involve:

The device should be removed from clinical service if the synchronization failure affects safe operation, alarm management, required documentation, or the facility's ability to use the equipment safely. When removal is required, label it Out of Service and send it for appropriate evaluation.

Service-level device issues should be:

After correction, verify the complete end-to-end path rather than relying only on a connected-network indication.

Knowing when the problem has moved beyond the anesthesia machine and into infrastructure or application support is proper troubleshooting.

Clinical Use Tip

When case-data synchronization fails, verify the entire path to the receiving clinical system and ensure staff have an approved alternate documentation workflow.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported that the Atlan A350 displayed the wrong network time and completed cases were not appearing in the expected downstream system."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the anesthesia machine connected to a network port that was not providing the required clinical network connectivity after a room connection change."

Resolution

What action was taken.

Example:
"Coordinated restoration of the correct network connection, verified time synchronization, confirmed successful case-data transfer to the receiving system, and returned the machine to service."

Helpful Details to Include (If Known)

Final Thought

Networked anesthesia-system problems require the same disciplined approach as hardware faults: protect clinical workflow, verify the physical connection first, isolate device versus infrastructure causes, avoid unauthorized configuration changes, and confirm the complete data path before closing the work order.

That is successful troubleshooting.