On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Addresses flow sensor calibration failures caused by sensor condition, installation, circuit setup, contamination, moisture, connections, or incompatible external components.
Step-by-Step Troubleshooting
1. Protect the Patient and Maintain Ventilation
If unreliable flow measurement affects active patient ventilation, monitoring, or alarms, move the patient to another verified ventilator or appropriate backup ventilation before troubleshooting.
Do not attempt calibration while a patient depends on the affected measurement system.
Expected outcome: The patient is safely supported and the ventilator is available for controlled testing. If safe patient support cannot be maintained, stop troubleshooting.
2. Confirm the Calibration Failure
Identify whether the reported problem is a failed calibration, unstable flow reading, inaccurate displayed volume, or sensor recognition problem.
Repeat the calibration only under the proper non-patient test conditions described by manufacturer documentation.
Expected outcome: The failure is reproducible and clearly associated with the flow measurement system. If calibration completes normally and subsequent measurements are verified, troubleshooting can stop.
3. Inspect the Flow Sensor
Remove and visually inspect the externally accessible flow sensor according to approved handling procedures.
Check for:
- Moisture
- Condensation
- Secretions or contamination
- Cracks
- Bent or damaged contacts
- Improper assembly
- Incorrect sensor type
Do not mechanically clean delicate sensing elements unless specifically permitted by manufacturer instructions.
Expected outcome: The sensor is clean, dry, compatible, and undamaged. If correcting the sensor condition resolves calibration, troubleshooting can stop.
4. Verify Sensor Installation and Orientation
Reinstall the flow sensor carefully and verify that it is fully seated and correctly oriented.
Inspect associated external connectors or cables for loose connections, contamination, or visible damage.
Expected outcome: The sensor is installed securely and recognized normally. If calibration succeeds after reseating, troubleshooting can stop after functional verification.
5. Inspect the Breathing Circuit
Verify the breathing circuit is properly assembled and free of major leaks, occlusions, water accumulation, or incorrect adapters that could interfere with calibration.
Confirm any required circuit components are installed in their intended positions.
Expected outcome: The circuit is correctly configured and does not create an external condition that prevents flow calibration. If correction restores calibration, troubleshooting can stop.
6. Allow Wet Components to Dry or Replace Them
If condensation or moisture is present, replace the affected flow sensor or circuit component with an approved dry component when appropriate.
Do not use improvised drying methods that could damage the sensor.
Expected outcome: Calibration is attempted with a clean, dry sensor and appropriate circuit. If calibration passes, troubleshooting can stop.
7. Substitute a Known-Good Flow Sensor
Install a compatible known-good sensor if available.
Repeat the calibration using the approved procedure.
Expected outcome: If the known-good sensor calibrates normally, the original sensor or its condition is implicated. Replace or process the original component according to facility and manufacturer requirements, then stop troubleshooting after successful verification.
8. Check for External Connection Problems
Inspect all accessible sensor interfaces for looseness, bent contacts, contamination, or cable strain.
Do not open the ventilator or troubleshoot internal electronics.
Expected outcome: External sensor interfaces are intact and stable. If correcting an external connection restores calibration, troubleshooting can stop.
9. Perform Final Functional Verification
After calibration succeeds, verify flow and volume measurements using appropriate approved test equipment and the applicable manufacturer procedure.
Confirm alarms and ventilation performance operate normally.
Expected outcome: Calibration completes successfully and measured performance is acceptable under the required return-to-service test. The ventilator may be returned to service when all required checks pass.
If the Problem Persists
If calibration continues to fail with a clean, dry, properly seated known-good compatible flow sensor and correct breathing-circuit setup, common external causes have been ruled out.
The remaining cause may involve an internal sensing interface, pneumatic pathway, calibration function, electronics, or other service-level problem.
The ventilator should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
- Evaluated using appropriate Drager documentation and approved test equipment
- Repaired or configured only by qualified personnel
Do not return the ventilator to service with unresolved or unreliable flow measurement.
Knowing when to stop external troubleshooting and escalate is proper troubleshooting.
Clinical Use Tip
Reliable neonatal flow measurement is critical for displayed volumes, alarms, and ventilator performance; do not accept intermittent sensor operation.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Respiratory Therapy reported that the Babylog VN600 would not complete flow sensor calibration."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found condensation present in the installed flow sensor."
Resolution
What action was taken.
Example:
"The wet sensor was replaced with a clean approved sensor, calibration completed successfully, and flow and ventilation performance were verified before return to service."
Helpful Details to Include (If Known)
- Exact calibration message
- Sensor type
- Sensor condition
- Presence of condensation or contamination
- Sensor seating and orientation
- Circuit configuration
- Known-good sensor substitution
- External connector condition
- Calibration results
- Flow or volume verification results
- Final device status
- Exact calibration message
- Sensor type
- Sensor condition
- Presence of condensation or contamination
- Sensor seating and orientation
- Circuit configuration
- Known-good sensor substitution
- External connector condition
- Calibration results
- Flow or volume verification results
- Final device status
Final Thought
Protect the patient first, evaluate sensor condition and setup before assuming an internal failure, ve### Helpful Details to Include (If Known)