On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Addresses failed oxygen calibration or unexpected FiO2 readings caused by gas supply, setup, sensor, circuit, connection, or configuration problems.
Step-by-Step Troubleshooting
1. Protect the Patient and Maintain Oxygen Delivery
If oxygen delivery or displayed FiO2 is questionable during patient use, move the patient to another verified ventilator or alternate approved ventilation source.
Do not troubleshoot uncertain oxygen delivery while a patient depends on the device.
Expected outcome: Reliable ventilation and oxygen delivery are maintained independently of the affected Babylog VN800.
2. Confirm the Reported Condition
Record the exact calibration failure, oxygen-related message, or discrepancy reported by staff.
Determine whether the problem involves failure to calibrate, an unexpected displayed FiO2, unstable readings, or disagreement with an independent analyzer.
Expected outcome: The oxygen-related symptom is clearly defined and repeatable.
3. Verify Oxygen Supply
Confirm the oxygen supply hose is fully connected and undamaged. Verify the source is available and that the connection has not been swapped, loosened, kinked, or restricted.
Use a known-good source or approved test method when necessary to distinguish ventilator issues from facility gas-supply problems.
Expected outcome: A reliable oxygen source is connected. If restoring the supply resolves the problem, complete calibration and functional verification before stopping.
4. Inspect Gas Hoses and External Connections
Inspect oxygen and air hoses, fittings, and accessible connection points for:
- Damage
- Loose fittings
- Kinks
- Contamination
- Incorrect connection
- Obvious leakage
Expected outcome: External gas connections are correct, secure, and unobstructed.
5. Verify Circuit and Test Setup
Inspect the patient circuit and test configuration. Confirm accessories, humidification components, filters, and sampling or measurement devices are installed correctly and are not introducing a leak or misleading measurement.
Expected outcome: The test setup is suitable for evaluating delivered oxygen concentration.
6. Perform the Approved Oxygen Calibration
With the ventilator removed from patient use, perform the normal approved oxygen calibration applicable to the installed configuration.
Do not repeatedly force calibration if the device consistently reports failure.
Expected outcome: Calibration completes normally. If it passes and subsequent FiO2 verification is satisfactory, troubleshooting can stop.
7. Compare With an Independent Oxygen Analyzer
When incorrect FiO2 is suspected, use an appropriate calibrated oxygen analyzer and approved test setup to compare delivered oxygen against the ventilator's displayed value.
Avoid assuming the ventilator is incorrect until the independent measurement setup has also been verified.
Expected outcome: Delivered and displayed oxygen values behave consistently with the selected test conditions. Significant unexplained disagreement requires further evaluation.
8. Check Accessible Sensor or Consumable Components
If the installed configuration includes an externally serviceable oxygen-sensing component, inspect it for age-related deterioration, contamination, moisture, improper installation, or damage.
Replace only with an approved compatible component and only when permitted by service procedure.
Expected outcome: External oxygen-sensing components are correctly installed and serviceable.
9. Verify Configuration and Settings
Confirm the selected oxygen setting, ventilation mode, and accessible configuration correspond to the intended test.
Do not alter restricted calibration parameters or service-level configuration without authorized documentation.
Expected outcome: No incorrect setting or configuration explains the reported discrepancy.
10. Complete Final Verification or Escalate
After correction, verify oxygen calibration, delivered FiO2 using approved test equipment, normal ventilation, and associated alarms.
If oxygen delivery remains unstable or cannot be verified, remove the ventilator from service.
Expected outcome: Oxygen delivery is verified and stable, allowing troubleshooting to stop, or the ventilator is escalated appropriately.
If the Problem Persists
External gas supply, hoses, circuit setup, calibration procedure, external sensing components, and accessible configuration have been checked. Remaining possibilities may involve internal gas-mixing, sensing, pneumatic, control, or calibration systems.
The ventilator should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
- Evaluated using appropriate Drager documentation and approved gas-analysis equipment
- Repaired or configured only by qualified personnel
After repair, verify oxygen delivery through the intended operating range according to applicable manufacturer and facility procedures before return to service.
Knowing when oxygen delivery cannot be reliably verified and stopping clinical use is proper troubleshooting.
Clinical Use Tip
When delivered oxygen is in doubt, maintain patient ventilation with another verified device rather than relying solely on the ventilator's displayed FiO2.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported the Babylog VN800 displayed an unexpected FiO2 and oxygen calibration would not complete."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found the oxygen supply hose incompletely connected at the ventilator inlet."
Resolution
What action was taken.
Example:
"The oxygen hose was securely reconnected, calibration completed successfully, and delivered FiO2 was verified with an approved oxygen analyzer."
Helpful Details to Include (If Known)
- Exact oxygen-related message
- Selected FiO2
- Displayed FiO2
- Independent analyzer reading
- Oxygen source tested
- Gas hose condition
- Circuit configuration
- Sensor condition
- Calibration result
- Results before and after correction
- Final device status
- Exact oxygen-related message
- Selected FiO2
- Displayed FiO2
- Independent analyzer reading
- Oxygen source tested
- Gas hose condition
- Circuit configuration
- Sensor condition
- Calibration result
- Results before and after correction
- Final device status
Final Thought
Questionable oxygen delivery requires immediate patient protection. Verify supply, hoses, circuit setup, sensing, calibration, and independent measurement before assuming int### Helpful Details to Include (If Known)