Drager Evita V500

Alarm: Airway obstructed? possible airway blockage or resistance causing reduced ventilation

On this page

Asset Type

Ventilator

Manufacturer

Drager

Model

Evita V500

What This Guide Helps With

Airway obstructed? alarm indicating possible blockage, high resistance, or low airflow impacting effective ventilation delivery to the patient.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Patient Airway (Most Critical Step)

Inspect Breathing Circuit for Obstruction

Check Tube or Mask Position and Integrity

Evaluate Ventilation Settings and Patient Interaction

Swap External Components if Needed

If the Problem Persists

If all external causes (airway, circuit, settings, accessories) have been ruled out and the alarm continues:

Clinical Use Tip

Never troubleshoot this alarm on an actively dependent patient without securing alternate ventilation first. Move the patient to a backup device before continuing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ventilator alarming 'Airway obstructed?' with low delivered volume during use."

Cause

What was observed during troubleshooting.

Example:
"Breathing circuit found kinked causing airflow restriction."

Resolution

What action was taken.

Example:
"Straightened/replaced circuit and verified normal ventilation."

Helpful Details to Include (If Known)

Final Thought

Airway obstruction alarms must always be taken seriously and addressed immediately. Start with the patient, then systematically rule out external causes before considering device failure. Proper escalation protects both the patient and equipment integrity.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide