Drager Evita V500

Alarm: Airway Pressure Low

On this page

Asset Type

Ventilator

Manufacturer

Drager

Model

Evita V500

What This Guide Helps With

Alarm: Airway Pressure Low indicating the ventilator cannot maintain set airway pressure, commonly due to leaks, disconnections, or failure within the breathing circuit, resulting in inadequate ventilation.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check Patient Circuit Connections

4. Check for Circuit Leaks or Damage

5. Verify Patient Interface (ET Tube or Mask)

6. Check Expiratory Valve Assembly

7. Review Ventilator Settings

8. Perform System Check

If the Problem Persists

If all external causes have been ruled out and the alarm continues, this indicates a likely internal issue such as a pressure sensor fault or internal leak.

Clinical Use Tip

Never troubleshoot a ventilator while actively supporting a patient. Always move the patient to a backup device before performing adjustments or troubleshooting steps.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ventilator alarming 'Airway Pressure Low' with reduced volume delivery during use."

Cause

What was observed during troubleshooting.

Example:
"Leak identified at loose circuit connection near humidifier."

Resolution

What action was taken.

Example:
"Secured connection and verified normal airway pressure restored."

Helpful Details to Include (If Known)

Final Thought

Low airway pressure alarms should be treated as potential ventilation failure. Always prioritize patient safety and check simple external causes like leaks and connections before escalating.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide