Drager Evita V500

Alarm: Alarm limit not confirmed

On this page

Asset Type

Ventilator

Manufacturer

Drager

Model

Evita V500

What This Guide Helps With

Alarm: Alarm limit not confirmed indicating that alarm parameter changes have not been accepted, which can prevent proper alarm monitoring and compromise safe ventilation operation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Confirm Alarm Limit Changes (Most Common Cause)

Check for Incomplete User Input or Timeout

Verify Control Interface Functionality

Power Cycle if Alarm Persists (Non-Patient Use Only)

If the Problem Persists

If the alarm continues after confirming settings and verifying interface function, external and user-input causes have been ruled out. Remove the device from service, label it as Out of Service, and send for repair or further evaluation.

Clinical Use Tip

Never troubleshoot on an active patient. Always confirm alarm limits immediately after making changes to ensure proper patient safety monitoring.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ventilator displaying 'Alarm limit not confirmed' alarm after alarm setting adjustments; alarm would not clear."

Cause

What was observed during troubleshooting.

Example:
"Alarm limits were changed but not confirmed through the control interface, leaving the system in a pending state."

Resolution

What action was taken.

Example:
"Re-entered alarm settings, confirmed all limits properly; alarm cleared. Verified normal operation and returned device to service."

Helpful Details to Include (If Known)

Final Thought

Alarm confirmation issues are typically related to workflow or user input rather than device failure. Always verify and confirm changes to maintain proper alarm function and ensure patient safety.

That is successful troubleshooting.