Drager Evita V500

Alarm: Ambient pressure sensor?

On this page

Asset Type

Ventilator

Manufacturer

Drager

Model

Evita V500

What This Guide Helps With

This guide helps troubleshoot the “Ambient pressure sensor?” alarm, which indicates inaccurate or unreliable ambient pressure readings that can affect ventilation accuracy, calibration, and device checks.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Altitude Setting (Most Common Cause)

Perform Device/System Check

Inspect Environmental Conditions

Power Cycle the Ventilator

Isolate as Internal Sensor Failure

If the Problem Persists

If all external configuration and environmental causes have been ruled out, the issue is likely an internal sensor failure.

Clinical Use Tip

Never troubleshoot a ventilator on an active patient. Always move the patient to a backup device before investigating alarms to ensure safe and uninterrupted ventilation.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ventilator displaying ‘Ambient pressure sensor?’ alarm during startup and failing device check."

Cause

What was observed during troubleshooting.

Example:
"Altitude setting verified correct; alarm persisted after power cycle and system test, indicating likely ambient pressure sensor failure."

Resolution

What action was taken.

Example:
"Removed ventilator from service, labeled Out of Service, and sent for repair and calibration."

Helpful Details to Include (If Known)

Final Thought

Ambient pressure sensor issues directly affect how the ventilator interprets and delivers pressure. Always verify configuration first, but escalate quickly if accuracy cannot be confirmed.

That is successful troubleshooting.