Drager Evita V500

Alarm: O2% low Measured oxygen concentration below set value

On this page

Asset Type

Ventilator

Manufacturer

Drager

Model

Evita V500

What This Guide Helps With

This guide addresses the “O2% low” alarm indicating the measured oxygen concentration is below the set value, which can result in inadequate oxygen delivery due to supply issues, sensor inaccuracies, or gas control problems.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check Oxygen Supply Source (Most Common Cause)

4. Inspect Gas Connections and Hoses

5. Verify Ventilator Settings

6. Calibrate the Oxygen Sensor

7. Run Extended System Test (EST)

8. Isolate Sensor vs Device Issue

If the Problem Persists

If all external causes, calibration, and sensor replacement have been ruled out, internal oxygen control or measurement components may be failing.

Clinical Use Tip

Never troubleshoot a ventilator with unresolved oxygen alarms while connected to a patient. Always transition the patient to a safe, reliable ventilation source first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ventilator alarming O2% low with measured oxygen below set FiO2 during patient use."

Cause

What was observed during troubleshooting.

Example:
"Low wall oxygen pressure and unstable sensor readings observed during calibration."

Resolution

What action was taken.

Example:
"Restored oxygen supply, calibrated sensor, and verified proper oxygen delivery. Replaced sensor if required."

Helpful Details to Include (If Known)

Final Thought

Low oxygen alarms are critical and should always be addressed immediately. Start with external supply and connections, verify sensor accuracy, and escalate when necessary to ensure safe patient care.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide