Drager Evita V500

Alarm: PCOMP not reached target pressure not achieved due to leak or incorrect tube settings

On this page

Asset Type

Ventilator

Manufacturer

Drager

Model

Evita V500

What This Guide Helps With

Alarm: PCOMP not reached indicating target pressure is not achieved, leading to inadequate ventilation due to leaks, disconnections, or incorrect tube configuration.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check for Circuit Leaks (Most Common Cause)

4. Verify Artificial Airway Integrity

5. Confirm Tube Type and I.D. Settings

6. Check Ventilator Settings

7. Inspect Expiratory Valve and Components

8. Perform a System Test (If Available)

If the Problem Persists

If all external causes such as leaks, settings, airway issues, and valve problems are ruled out and the alarm continues:

Clinical Use Tip

Never troubleshoot this alarm on an active patient. Always transfer the patient to a backup ventilator before evaluating the device.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ventilator alarming PCOMP not reached with low delivered pressure during use."

Cause

What was observed during troubleshooting.

Example:
"Leak identified at patient circuit connection preventing pressure buildup."

Resolution

What action was taken.

Example:
"Re-secured circuit connection, verified no leaks, and confirmed normal pressure delivery."

Helpful Details to Include (If Known)

Final Thought

Pressure delivery alarms should always be treated as real ventilation failures. Start with leaks and airway integrity before escalating. Proper troubleshooting and documentation ensure patient safety and effective resolution.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide