Drager Evita V500

Alarm: PVENT high Inspiratory pressure exceeds safe threshold

On this page

Asset Type

Ventilator

Manufacturer

Drager

Model

Evita V500

What This Guide Helps With

This guide addresses the PVENT high alarm, indicating excessive inspiratory pressure that may lead to lung injury, ineffective ventilation, or circuit-related issues.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Breathing Circuit for Obstruction

Inspect Patient Interface (ET Tube / Mask)

Review Ventilation Settings

Check Expiratory Valve Function

Run System Self-Test (SST)

Run Extended System Test (EST)

If the Problem Persists

If all external causes have been ruled out and the alarm continues:

Clinical Use Tip

Never troubleshoot a ventilator on an active patient. Always transfer the patient to a backup device before continuing troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ventilator displaying PVENT high alarm with elevated inspiratory pressure during use."

Cause

What was observed during troubleshooting.

Example:
"No circuit obstruction or airway issue identified; pressure remained elevated, suggesting possible internal sensor or valve issue."

Resolution

What action was taken.

Example:
"Removed ventilator from service and sent for repair. Backup ventilator provided."

Helpful Details to Include (If Known)

Final Thought

High inspiratory pressure alarms require immediate attention. Always rule out external causes first, and escalate quickly if pressure control cannot be verified. Proper documentation and patient safety must remain the priority.

That is successful troubleshooting.