Drager Evita V300 Ventilator

Flow Sensor Error

On this page

Asset Type

Ventilator

Manufacturer

Drager

Model

Evita V300

What This Guide Helps With

This guide addresses flow sensor error messages on the Evita V300 ventilator. These alarms typically indicate that the ventilator is not receiving accurate airflow measurements from the flow sensor. Common causes include moisture buildup, contamination, loose connections, or improperly seated components. This guide focuses on verifying simple, external issues before assuming internal sensor failure.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Check the Alarm Message on the Display

Inspect the Flow Sensor Location

Check for Moisture or Condensation

Inspect the Breathing Circuit

Reseat the Flow Sensor

Perform a Pre-Use Check

Try a Known-Good Flow Sensor

Inspect the Sensor Port on the Ventilator

If the Problem Persists

If the flow sensor alarm continues after:

The issue may involve:

At this point, the ventilator should be:

Recognizing when the issue is likely internal and stopping further troubleshooting is proper Clinical Engineering practice.

Clinical Use Tip

Never troubleshoot a ventilator while it is actively supporting a patient. Always ensure the patient is safely moved to another ventilator or manual ventilation before removing sensors or circuit components.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Respiratory reported Evita V300 displaying a flow sensor error alarm during setup.”

Cause

What was observed during troubleshooting.

Example:
“Flow sensor found with visible condensation and not fully seated in the sensor housing.”

Resolution

What action was taken.

Example:
“Sensor removed, dried, and reseated. Pre-use check performed and ventilator passed all tests. Ventilator returned to service.”

Helpful Details to Include (If Known)

Final Thought

Ventilator flow sensors are sensitive components that can be affected by moisture, contamination, or improper installation. Starting with simple external checks such as sensor placement, circuit condition, and moisture inspection often resolves the issue quickly. When those checks do not correct the problem, escalating the device for further evaluation protects both the equipment and patient safety.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide