Drager Evita V500

Alarm: Compliance limited VT delivered tidal volume limited due to compliance compensation limits

On this page

Asset Type

Ventilator

Manufacturer

Drager

Model

Evita V500

What This Guide Helps With

This guide addresses the “Compliance limited VT” alarm, which indicates that the ventilator is unable to deliver the set tidal volume due to circuit compliance or compensation limits. This can result in reduced effective ventilation and inaccurate volume delivery to the patient.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Breathing Circuit Connections

Inspect Circuit for Excess Compliance or Expansion

Check for Leaks in Circuit or Interfaces

Verify Ventilator Settings

Run System Test (SST)

Swap Circuit if Needed

If the Problem Persists

If all external causes such as circuit issues, leaks, and settings have been ruled out, an internal compliance measurement or compensation failure is likely.

Clinical Use Tip

Never troubleshoot compliance or volume delivery alarms on an active patient. Always transfer the patient to a backup ventilator before making adjustments.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ventilator alarming 'Compliance limited VT' with low delivered tidal volume during use."

Cause

What was observed during troubleshooting.

Example:
"Excess circuit compliance due to loose connection and worn tubing (or) no external issue found, suspected internal compliance compensation limitation."

Resolution

What action was taken.

Example:
"Reseated connections and replaced breathing circuit; verified proper VT delivery and alarm cleared (or) removed from service after failed SST and persistent alarm."

Helpful Details to Include (If Known)

Final Thought

Compliance-related alarms reflect real limitations in volume delivery. Always start with external components like circuits and connections before escalating to internal faults. Proper troubleshooting protects both patient safety and device reliability.

That is successful troubleshooting.