Drager Evita V600

Apnea, Low Minute Volume, or Low Tidal Volume Alarm

On this page

Asset Type

Ventilator

Manufacturer

Drager

Model

Evita V600

What This Guide Helps With

Addresses low-volume or apnea alarms caused by disconnections, leaks, circuit problems, sensing issues, settings, patient conditions, or ventilator faults.

Step-by-Step Troubleshooting

1. Protect the Patient and Confirm Effective Ventilation

Treat apnea and low-volume alarms as potentially clinical emergencies. Clinical staff should immediately assess the patient, airway, chest movement, and actual ventilation.

If effective ventilation is uncertain, use an appropriate backup method or another verified ventilator.

Expected outcome: Patient ventilation is maintained independently of the troubleshooting process.

2. Confirm the Exact Alarm

Identify whether the reported condition is apnea, low minute volume, low tidal volume, or a combination. Record the exact displayed alarm and when it occurs.

Expected outcome: The specific reported condition and its timing are documented.

3. Check for Circuit Disconnection

Trace the entire breathing circuit from ventilator to patient connection. Look for loose fittings, disconnected tubing, partially separated components, or an open sampling or accessory port.

Expected outcome: The breathing circuit is fully connected. If reconnecting a loose component restores ventilation and alarms clear, proceed to final verification.

4. Check for Circuit or Airway Leaks

Inspect the circuit, humidification components, tubing, adapters, and accessible connections for cracks, loose fittings, or other leakage.

Patient airway and interface leakage should be evaluated by clinical staff.

Expected outcome: Significant external leaks are identified and corrected.

5. Inspect Flow-Sensing Components

Verify correct seating and condition of accessible flow measurement components. Look for contamination, moisture, or improper installation.

Expected outcome: Flow sensing components are correctly installed and suitable for operation.

6. Verify Ventilation Settings

Confirm that ventilation mode, volume or pressure settings, alarm limits, and apnea-related settings correspond to the intended clinical plan.

Clinical settings should be confirmed with the responsible respiratory or medical team rather than changed solely to silence an alarm.

Expected outcome: The ventilator configuration is appropriate for the intended use.

7. Test the Ventilator With a Test Lung

Remove the ventilator from patient service before technical testing. Connect a suitable test lung and determine whether the reported low-volume or apnea condition can be reproduced.

Expected outcome: Stable volume delivery and monitoring occur with the test lung. If the alarm occurs only on the patient, clinical factors should be evaluated.

8. Substitute Known-Good External Components

If the problem persists off patient, replace questionable external components such as the breathing circuit, flow sensor, or relevant accessory with known-good approved components.

Expected outcome: The fault clears when the defective external component is removed from the system.

9. Verify Delivered and Measured Ventilation

Use appropriate approved test equipment to evaluate delivered and measured ventilation following correction.

Expected outcome: Ventilation measurements are stable and alarm behavior is appropriate. Troubleshooting can stop when required verification passes.

10. Escalate Persistent Low-Volume or Apnea Alarms

If alarms continue during controlled testing with a verified circuit and appropriate external components, stop external troubleshooting.

Expected outcome: The device remains out of service for qualified technical evaluation.

If the Problem Persists

External disconnections, leaks, circuit components, flow sensing, and settings have been evaluated. Remaining possibilities may involve internal pneumatic delivery, flow or pressure sensing, valve control, electronics, software, or calibration.

The ventilator should be:

Complete appropriate ventilation-performance, alarm, sensing, and system testing after repair. Stopping when safe external troubleshooting has been exhausted is proper troubleshooting.

Clinical Use Tip

Never assume a low-volume or apnea alarm is caused by the ventilator until the patient, airway, and breathing circuit have been immediately assessed.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Respiratory Therapy reported repeated low tidal volume alarms from the Evita V600."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found a loose connection in the external breathing circuit that created a significant leak."

Resolution

What action was taken.

Example:
"The circuit connection was secured, ventilation was verified with a test lung, and appropriate volume measurement and alarm operation were confirmed before return to service."

Helpful Details to Include (If Known)

Final Thought

Apnea and low-volume alarms require patient assessment first, followed by logical examination of the breathing circuit, leaks, sensing components, and configuration. Equipment should only return to service after the reported condition is resolved and ventilation performance has been independently verified.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide