Drager Evita V600

Flow Sensor Calibration Failure

On this page

Asset Type

Ventilator

Manufacturer

Drager

Model

Evita V600

What This Guide Helps With

Addresses flow sensor calibration failures caused by contamination, moisture, installation problems, damaged sensors, breathing-system setup, or unresolved service-level faults.

Step-by-Step Troubleshooting

1. Protect the Patient and Maintain Ventilation

Do not troubleshoot unreliable flow measurement while a patient depends on the affected ventilator. Transfer the patient to another verified ventilator or appropriate backup ventilation if the failure affects safe monitoring or ventilation.

Expected outcome: Patient ventilation remains supported independently of the device being evaluated.

2. Confirm the Calibration Failure

Identify whether calibration will not begin, stops before completion, or completes unsuccessfully. Record any displayed message exactly.

Confirm whether the problem occurs every time or began after circuit cleaning, sensor replacement, or breathing-system reassembly.

Expected outcome: The failure is reproduced and its circumstances are documented.

3. Inspect the Flow Sensor

Remove the flow sensor only according to normal approved handling procedures. Inspect it for visible contamination, moisture, residue, damage, incorrect assembly, or deformation.

Do not attempt component-level repair of the sensor.

Expected outcome: The sensor is clean, dry when required, physically intact, and suitable for testing. If correcting an obvious external issue restores calibration, proceed to final verification.

4. Verify Sensor Installation

Reinstall the sensor carefully and confirm that it is fully seated and oriented correctly in its intended location.

Inspect accessible mating surfaces and connections for contamination, damage, or incomplete engagement.

Expected outcome: The sensor is correctly installed and securely seated. If calibration succeeds after reseating, troubleshooting can stop after final verification.

5. Inspect the Breathing System

Check the circuit, expiratory valve, filters, humidification components, and related accessories for incorrect assembly, leaks, moisture, or obstruction that could interfere with calibration.

Expected outcome: The breathing system is correctly assembled and free of obvious conditions that could affect flow measurement.

6. Check for Moisture or Condensation

Inspect the externally accessible flow path for accumulated moisture or condensation, especially following cleaning or humidified operation.

Correct moisture problems using approved handling procedures and ensure reusable components are prepared according to facility and manufacturer requirements.

Expected outcome: Excess moisture is removed and components are in an appropriate condition for calibration.

7. Substitute a Known-Good Approved Flow Sensor

If available, install a known-good compatible sensor and repeat the calibration.

Do not condemn the ventilator based solely on one questionable sensor.

Expected outcome: Calibration succeeds with the known-good sensor, identifying the original sensor as the likely external cause. Replace the affected sensor and proceed to verification.

8. Compare Against a Known-Good Setup

If calibration still fails, compare the breathing-system installation and removable components with another correctly configured Evita V600.

Correct only verified setup differences.

Expected outcome: External configuration differences are eliminated as a cause.

9. Perform Functional Verification

After successful calibration, operate the ventilator using appropriate test equipment or a test lung. Verify that displayed flow-related measurements respond plausibly and that ventilation is stable.

Expected outcome: Calibration completes successfully and flow measurement behaves normally. Troubleshooting can stop and the ventilator may be returned to service according to policy.

10. Escalate Persistent Calibration Failure

If calibration fails with a known-good sensor and correct external setup, discontinue external troubleshooting.

Expected outcome: The device remains out of clinical service until evaluated by qualified service personnel.

If the Problem Persists

External sensor condition, installation, circuit setup, and removable components have been ruled out. Remaining causes may involve internal sensing, pneumatic, electronic, calibration, or service-level problems.

The ventilator should be:

After repair, complete required calibration and ventilator performance testing before return to service. Stopping when external troubleshooting no longer provides a safe path forward is proper troubleshooting.

Clinical Use Tip

Do not rely on a ventilator with unresolved flow-measurement or calibration problems for patient ventilation or volume monitoring.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Respiratory Therapy reported that the Evita V600 repeatedly failed flow sensor calibration."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found moisture contamination on the removable flow sensor following breathing-circuit use."

Resolution

What action was taken.

Example:
"The affected sensor was replaced with an approved known-good sensor, calibration completed successfully, and ventilation was verified with a test lung."

Helpful Details to Include (If Known)

Final Thought

Flow calibration problems should be addressed by first protecting the patient and then verifying the sensor, installation, breathing system, and moisture conditions before assuming an internal fault. Successful calibration must be followed by functional verification and clear CCR documentation.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide