Drager Evita V600

High Airway Pressure Alarm

On this page

Asset Type

Ventilator

Manufacturer

Drager

Model

Evita V600

What This Guide Helps With

Addresses persistent high airway pressure alarms caused by circuit obstruction, tubing problems, accessories, settings, patient-related resistance, or ventilator malfunction.

Step-by-Step Troubleshooting

1. Protect the Patient and Assess the Clinical Situation

A high airway pressure alarm can indicate an immediate patient or airway problem. Clinical staff should assess the patient, airway, and ventilation first.

Do not troubleshoot the ventilator while a patient depends on unreliable ventilation. Exchange the device if an equipment fault is suspected and cannot be immediately ruled out.

Expected outcome: The patient's ventilation and airway are safely managed before equipment troubleshooting begins.

2. Confirm the Alarm Condition

Record the exact alarm and determine whether it occurs continuously, during inspiration, only at particular settings, or only with a specific circuit or patient setup.

Expected outcome: The circumstances that trigger the alarm are clearly identified.

3. Inspect the Breathing Circuit

Look for kinked tubing, compressed tubing, blocked filters, closed clamps, pooled condensate, obstructed connectors, or incorrectly installed accessories.

Trace the entire inspiratory and expiratory pathway.

Expected outcome: The breathing circuit is open and correctly assembled. If removing an external obstruction eliminates the alarm, proceed to final verification.

4. Inspect the Expiratory Path

Verify that the expiratory valve and associated external components are properly installed and unobstructed.

Check for moisture, secretions, contamination, or incorrect seating that could increase expiratory resistance.

Expected outcome: Exhaled gas has an unrestricted pathway through correctly installed external components.

5. Check Filters, Humidification, and Accessories

Inspect filters, humidifiers, adapters, nebulizer components, and other added accessories for obstruction, saturation, incorrect orientation, or excessive resistance.

Substitute known-good components when appropriate.

Expected outcome: No external accessory is producing abnormal resistance. If replacing an obstructed accessory resolves the alarm, troubleshooting can stop after verification.

6. Verify Ventilator Settings

Confirm that pressure-related limits and ventilation settings match the intended clinical configuration. Clinical settings should be evaluated in coordination with qualified clinical staff.

Do not arbitrarily increase an alarm threshold to suppress a valid alarm.

Expected outcome: Settings are appropriate and the alarm is not being caused by an unintended configuration.

7. Test the Ventilator Off Patient

After removing the ventilator from patient use, connect an appropriate test lung and reproduce the operating conditions as closely as practical.

Observe whether the high-pressure alarm persists without the patient.

Expected outcome: If the alarm does not occur with a test lung, patient, airway, or patient-specific circuit conditions require clinical evaluation. If it persists, continue equipment troubleshooting.

8. Substitute the Breathing Circuit

Install a known-good approved circuit and appropriate external accessories.

Expected outcome: If the alarm resolves with the replacement circuit, the original circuit or accessory set was the likely cause.

9. Perform Final Functional Verification

Operate the ventilator over appropriate test conditions using a test lung and approved test equipment. Confirm appropriate pressure behavior and alarm response.

Expected outcome: Ventilation is stable, pressures behave as expected, and alarm operation is functional. Troubleshooting can stop.

10. Escalate Persistent Equipment-Related High Pressure

If the high-pressure condition persists using a verified test lung, known-good circuit, correct external components, and appropriate settings, stop external troubleshooting.

Expected outcome: The ventilator is removed from service for further evaluation.

If the Problem Persists

External circuit obstruction, accessories, settings, and patient-related conditions have been separated from the equipment problem. Remaining causes may involve internal pneumatic control, valves, pressure sensing, electronics, or service-level calibration.

The ventilator should be:

After service, perform required ventilation, pressure, alarm, and system verification before return to service. Knowing when persistent pressure abnormalities require escalation is proper troubleshooting.

Clinical Use Tip

Always distinguish a true patient or airway obstruction from an equipment problem before continuing technical troubleshooting of a high-pressure alarm.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Respiratory Therapy reported repeated high airway pressure alarms from the Evita V600."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found a saturated external breathing-circuit filter creating excessive resistance."

Resolution

What action was taken.

Example:
"The affected filter was replaced, ventilation was tested with a test lung, and normal pressure and alarm behavior were verified before return to service."

Helpful Details to Include (If Known)

Final Thought

High airway pressure alarms require immediate attention to the patient before technical troubleshooting. Once patient factors are separated from the equipment, systematic inspection of the circuit, accessories, expiratory path, and settings can identify common external causes without prematurely assuming an internal ventilator fault.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide