On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Addresses charging problems or unexpectedly short battery runtime associated with AC power, connections, battery condition, storage history, or charging-system faults.
Step-by-Step Troubleshooting
1. Protect the Patient and Plan for Loss of Power
A ventilator with uncertain battery performance should not be relied upon for transport or continued ventilation during possible AC interruption. If the device is supporting a patient, maintain it on verified AC power and arrange another ventilator before troubleshooting if battery reliability is questionable.
Expected outcome: Patient ventilation does not depend on an unreliable battery.
2. Confirm the Exact Battery Complaint
Determine whether the complaint is:
- Battery does not charge
- Charge level does not increase
- Battery depletes rapidly
- Unit shuts down sooner than expected
- Battery is not recognized
- Problem occurs only during transport
Record the battery indication and operating condition.
Expected outcome: The reported battery behavior is clearly characterized. If normal behavior is confirmed after a documented incomplete-charge condition, complete verification and troubleshooting can stop.
3. Verify AC Power
Confirm the ventilator is connected to a verified AC source and that its external power connection is fully seated. Check the power cord for damage, looseness, or signs of overheating.
If the outlet is questionable, verify it using an appropriate facility-approved method.
Expected outcome: Reliable AC power reaches the ventilator. If restoring AC power starts normal charging and subsequent battery testing passes, troubleshooting can stop.
4. Observe Charging Indications
With verified AC connected, observe the ventilator's normal user-accessible battery and charging indications. Note whether the battery is detected and whether charge status changes appropriately over time.
Do not assume a battery has failed simply because it was initially at a low state of charge.
Expected outcome: The battery is recognized and indicates charging. If normal charge behavior and subsequent runtime are demonstrated, troubleshooting can stop.
5. Inspect External Power Components
Inspect the accessible AC cord, plug, strain relief, and connector for physical damage, contamination, or looseness. Replace damaged detachable power components with compatible approved replacements.
Expected outcome: External power components are intact and secure. If replacement restores normal charging, continue through functional battery verification.
6. Review Battery Use and Storage History
Determine whether the ventilator was recently stored, left disconnected from AC, repeatedly used for transport, or returned with a depleted battery.
This helps distinguish inadequate charging opportunity from an actual battery-performance problem.
Expected outcome: Recent use history is understood. If the battery simply lacked adequate charging opportunity and subsequently passes testing, troubleshooting can stop.
7. Inspect Accessible Battery Condition
If the battery is externally accessible according to normal service practice, inspect for obvious swelling, leakage, cracking, loose fit, contamination, or physical damage.
Do not use a visibly damaged battery.
Expected outcome: No unsafe physical battery condition is present. Any damaged battery is removed from service and handled according to facility battery procedures.
8. Test Battery Performance
After appropriate charging, perform a battery performance or runtime evaluation using the applicable manufacturer or facility-approved method. Do not invent runtime thresholds.
Expected outcome: Battery performance meets approved acceptance criteria. If it passes, the device can proceed to final verification.
9. Verify AC-to-Battery Transition
When permitted by the approved test procedure and with the ventilator off the patient, verify that the device transitions appropriately between AC power and battery operation and that battery status indications remain normal.
Expected outcome: The ventilator continues operating normally during the controlled transition and returns to charging when AC is restored. Troubleshooting can stop if all criteria are met.
10. Escalate Persistent Charging or Runtime Failure
If verified AC is present but the battery is not recognized, does not charge, fails approved performance testing, or the ventilator behaves abnormally during power transition, remove it from service.
Expected outcome: The ventilator is held for qualified battery or charging-system evaluation.
If the Problem Persists
After AC source, power cord, charging opportunity, accessible battery condition, and approved battery testing have been evaluated, persistent failure may involve battery degradation, charging circuitry, power-management electronics, battery communication, or another service-level condition.
The ventilator should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
- Evaluated using appropriate manufacturer documentation and approved test equipment
- Repaired or configured only by qualified personnel
After corrective work, verify charging, battery performance, AC/battery transition, alarms, and normal ventilator operation before return to service. Knowing when battery reliability cannot be assured is proper troubleshooting.
Clinical Use Tip
Do not use an Evita V800 with questionable battery performance for patient transport where dependable backup power is required.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Respiratory Therapy reported the Evita V800 battery charge remained low and the unit had reduced runtime during transport preparation."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found the AC power cord was incompletely seated, preventing reliable charging while the ventilator was parked."
Resolution
What action was taken.
Example:
"Clinical Engineering secured the AC connection, confirmed normal charging, completed approved battery performance and power-transition testing, and returned the ventilator to service."
Helpful Details to Include (If Known)
- Reported battery symptom
- Battery status indication
- AC source tested
- Power cord condition
- Charging indication
- Recent storage or transport history
- Visible battery condition
- Approved performance-test result
- AC-to-battery transition result
- Final device status
Final Thought
Battery complaints require verification of the entire external charging path before the battery or ventilator is blamed. Protect the patient from unexpected power loss, test battery performance using approved criteria, and remove the ventilator from service whenever backup power cannot be trusted.
That is successful troubleshooting.