Drager Evita V800

Flow Sensor Calibration Failure

On this page

Asset Type

Ventilator

Manufacturer

Drager

Model

Evita V800

What This Guide Helps With

Addresses unsuccessful flow-sensor calibration caused by sensor installation, contamination, moisture, incompatible components, circuit setup, or other external conditions.

Step-by-Step Troubleshooting

1. Protect the Patient Before Troubleshooting

Do not perform disruptive flow-sensor troubleshooting while a patient depends on the affected ventilator. Transfer the patient to another verified ventilator or maintain ventilation using an appropriate backup method.

Expected outcome: The patient remains continuously and reliably ventilated. If dependable monitoring or ventilation cannot be maintained, remove the Evita V800 from clinical use.

2. Confirm the Calibration Failure

Verify the exact reported condition and determine whether the flow sensor fails calibration consistently. Record the displayed message and whether the failure occurred during startup, a system check, or a separate calibration attempt.

Expected outcome: The calibration failure is confirmed and clearly documented. If calibration now completes normally, perform functional verification and troubleshooting can stop.

3. Inspect the Flow Sensor Installation

Verify that the applicable flow sensor is:

Do not force connectors or modify the sensor.

Expected outcome: The flow sensor is correctly installed. If reseating it allows calibration to complete successfully, verify ventilation performance and troubleshooting can stop.

4. Inspect for Moisture or Contamination

Remove the ventilator from patient use before handling respiratory components. Inspect the flow sensor and surrounding accessible components for condensate, secretions, residue, debris, or contamination that could interfere with sensing.

Follow approved cleaning, drying, or replacement practices rather than attempting unauthorized repair.

Expected outcome: The sensing path is clean, dry, and unobstructed. If an approved cleaned, dried, or replacement sensor calibrates successfully, troubleshooting can stop after verification.

5. Inspect the Breathing Circuit

Check the circuit for improper connections, occlusions, excessive water accumulation, open ports, leaks, or components that could disturb flow during calibration.

Ensure the circuit is arranged as required for the check being performed.

Expected outcome: The breathing circuit is correctly assembled and free of obvious leaks or obstructions. If calibration succeeds after correcting the circuit, troubleshooting can stop.

6. Verify Expiratory Components

Check that the expiratory valve and other externally removable components associated with expiratory flow measurement are correctly installed and fully seated.

Inspect for obvious damage or contamination.

Expected outcome: Expiratory components are correctly installed and recognized. If correction restores successful flow calibration, repeat the applicable system check and troubleshooting can stop.

7. Test With a Known-Good Compatible Flow Sensor

If available, substitute a compatible known-good flow sensor. Change only the sensor so the test remains meaningful.

Expected outcome: Calibration succeeds with the known-good sensor or the sensor is ruled out as the cause. If the replacement calibrates normally, remove the suspect sensor from use and troubleshooting can stop after final verification.

8. Verify User-Accessible Configuration

Confirm the ventilator is configured for the installed patient circuit and sensor arrangement. Compare with a known-good Evita V800 setup if necessary.

Do not alter protected service parameters or access unauthorized service menus.

Expected outcome: User-accessible configuration matches the installed respiratory setup. If correcting configuration permits successful calibration, repeat functional checks and troubleshooting can stop.

9. Complete Functional Verification

After successful calibration, perform the applicable ventilator system check. Use appropriate approved test equipment or a test lung as required by facility procedure to confirm consistent flow and volume behavior and proper alarm function.

Expected outcome: Flow calibration completes successfully and the ventilator demonstrates normal measurement and ventilation behavior. The device can be returned to service according to facility procedure.

10. Escalate Persistent Calibration Failure

If calibration continues to fail with a known-good compatible sensor and correct circuit configuration, stop external troubleshooting.

Expected outcome: The ventilator remains out of clinical service pending qualified evaluation.

If the Problem Persists

Once sensor seating, cleanliness, circuit setup, expiratory components, compatible sensor substitution, and accessible configuration have been ruled out, the failure may involve an internal sensing, pneumatic, electronic, calibration, or service-level problem.

The ventilator should be:

Following repair, complete applicable calibration, performance, alarm, and return-to-service testing. Recognizing that a persistent calibration failure requires escalation is proper troubleshooting.

Clinical Use Tip

Do not rely on a ventilator for patient flow or volume monitoring when its required flow-sensor calibration cannot be completed successfully.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Respiratory Therapy reported the Evita V800 would not complete flow-sensor calibration during pre-use testing."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found moisture contamination on the removable flow sensor that interfered with calibration."

Resolution

What action was taken.

Example:
"Clinical Engineering replaced the affected flow sensor with a compatible known-good sensor, completed calibration and the system check successfully, verified ventilation operation, and returned the unit to service."

Helpful Details to Include (If Known)

Final Thought

Maintain patient ventilation first, then verify sensor installation, cleanliness, circuit setup, and compatible components before suspecting the ventilator itself. Successful calibration must be followed by functional verification and clear documentation, while persistent failures require escalation.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide